D-11049-2, Ver 11
Page 23 of 26
MAINTENANCE INSTRUCTIONS
This product requires little maintenance. It is still recommended that a trained technician check the product every 12 months.
For the sanitizing/disinfecting wipes, please refer to the manufacturer's cautions and instructions for use, storage, and disposal.
A
MPLITRODE
®,
V
IVO
A
MP
™,
AND
E
LECTRODE
C
LIPS
The Amplitrode®, VivoAmp™, and the electrode clips should be wiped with disinfecting wipes after each use.
ABR
T
RANSDUCER AND ELECTRODES
ER-3A-800/ER-3C-800
The ear tips of the ER-3A-800/ER-3C-800 Insert Earphones, and the snap electrode pads are both disposable and should be properly
disposed of after each use.
B71W Bone Conductor
The B71W and the headband should be disinfected with sanitizing wipes after each use.
C
LEANING THE
OAE
P
ROBE
For sanitary reasons and to maintain acoustic performance, the Probe must be kept clean of any visible particles, such as earwax,
dead skin, and any other debris.
1.
During cleaning of the Probe, hold the Probe tip down to avoid pushing the debris deeper into the sound canals of the Probe.
2.
Brush off any visible debris from the Probe surface with the mini-brush.
3.
Remove remaining particles from the Probe ditches with the all-purpose pick.
4.
Remove particles from the sound canals and vent with the wax removal tool. The vent (a tiny hole above the cable at the back
of the Probe) is to ensure the release excessive pressure from the occluded ear canal.
D
ISINFECTING THE
OAE
P
ROBE
1.
The Probe should be wiped with disinfecting wipes after each use. Hold the Probe tip down to avoid any ear wax or sanitizing
liquid from entering the microphone sound canals.
2.
Gently clean the Probe with a sanitizing wipe, and dispose of the wipe after use. Excessive force is not needed.
3.
Insert the Probe tightly into the tapered hole of the Probe Holder.
4.
Wipe off any excess liquid, and allow the Probe to dry for about an hour.
5.
Perform a Probe Cavity Check. If a problem is detected, contact Customer Support. The problem may be due to residual ear
wax on the microphone sound canals.
EC
OCH
G
G
OLD
E
LECTRODE
E
AR
T
IP
Properly disposed of the gold electrode (TIPtrode) ear tips after each use.
C
ARING FOR THE
V
IVO
L
INK
The VivoLink may be wiped with disinfecting wipes after each use.
C
ARING FOR THE
C
OMPUTER
See the computer manufacturer's User Manual.
D
ISPOSAL OF
E
LECTRONIC
E
QUIPMENT
(WEEE)
The equipment shall not be disposed of as unsorted municipal waste and shall be collected separately as electrical and electronic
equipment, as applicable, as specified by Directive 2012/19/EU of The European Parliament and The Council of the European Union
on Waste Electrical and Electronic Equipment (WEEE).