8
WHISPER-0
USER MANUAL
www.vicidbs.com
Version 1.03.0000
4.5. Process Diagram
Compressed
Air
Whisper-0
LC-MS
(or
other)
5. INSTALLATION
5.1. Transport
Transport the Whisper-0 without tipping it in the original packing.
5.2. Location
Install the Whisper-0 in an area with the following characteristics:
•
Minimum distance of 10 cm at the front and lateral sides to allow sufficient cooling
• An enclosed (indoor), dry room
• Avoid direct sunlight
• Keep away from heat sources
• Room with adequate aeration/ventilation
•
Connect a tube for waste humidity from filter “B”
5.3. Connecting the Whisper-0 to the Air Compressor
There are 2 configuration possibilities:
• If you use a high capacity compressor (like Scroll compressor) without buffer-tank between the Whisper-0
and the LC-MS, then turn the red tap on the back side of the generator as shown in
Figure 1
below.
• If you have a normal/standard compressor and a buffer-tank (min 100 lt) between the Whisper-0 and the
LC-MS then turn the red tap on the back side of the generator as shown in
Figure 2
below.
Figure 1
Figure 2