6
WHISPER-0
USER MANUAL
www.vicidbs.com
Version 1.03.0000
A special characteristic of the Whisper 0-Hybrid is the momentary stop of the compressor if the LC-MS is not
using nitrogen. This momentary stop is regulated by a pneumatic control valve pre-calibrated by VICI DBS to
lengthen the life span of the compressor and at the same time help save energy.
4.2. Spare Parts
PART NUMBER
DESCRIPTION
DB-WN2-001
Annual Filter Kit (3 filters)
DB-WN2-002
Pneumatic Valve
DB-WN2-003
Manometer
DB-WN2-005
10 liter tank
DB-WN2-006
Tube 10 m
DB-WN2-014
Valveless N2 small membrane
DB-WN2-015
Filter housing for A0035 element without fittings
DB-WN2-016
Filter housing for V0035 and S0035 element without fittings
4.3. Seperation Principle
A
F
S
H2O - H2
O2
N2
B
C
A
Compressed Air Input
B
Hollow Fiber Membrane
C
Nitrogen Output
Atmospheric air contains nitrogen (78.1%), oxygen (20.9%), argon (1%), carbon dioxide, water vapor, and
traces of inert gases. The compressed air (A) is forced through the hollow fiber membrane (B). The various
components of the air permeate the porous membrane walls.