Calibration 17
Reference Table 1
RANGE
RESISTOR
STANDARD
VALUE
WAITING TIME
BEFORE
PRESSING
KEYS
PRESS KEY
300mΩ
0Ω
5 sec …
SHIFT+TRIG
SHIFT+HOLD
100mΩ
3Ω
0Ω
3 sec …
SHIFT+TRIG
SHIFT+HOLD
1Ω
30Ω
0Ω
3 sec …
SHIFT+TRIG
SHIFT+HOLD
10Ω
300Ω
0Ω
3 sec …
SHIFT+TRIG
SHIFT+HOLD
100Ω
3KΩ
0Ω
3 sec …
SHIFT+TRIG
SHIFT+HOLD
1kΩ
30KΩ
0Ω
3 sec …
SHIFT+TRIG
SHIFT+HOLD
10kΩ
300KΩ
0Ω
3 sec …
SHIFT+TRIG
SHIFT+HOLD
100kΩ
3MΩ
0Ω
5 sec …
SHIFT+TRIG
SHIFT+HOLD
1MΩ
30MΩ
0Ω
10 sec …
SHIFT+TRIG
SHIFT+HOLD
10MΩ
Note:
DO NOT
enter the adjustment mode if the Resistor Standard and
a certified technician are not available.