This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received,
including interference that may cause undesired
operation.
1. The Blood Pressure Monitor needs special
precautions regarding EMC and needs to be
installed and put into service according to the
EMC information provided in the ACCOMPANYING
DOCUMENTS
2. Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-
talkies can affect this equipment and should be
kept at least a distance d = 3,3 m away from the
equipment.
(Note. As indicated in Table 6 of IEC 60601-1-2:2007 for
ME EQUIPMENT, a typical cell phone with a maximum
output power of 2 W yields d = 3,3 m at an IMMUNITY
LEVEL of 3 V/m)
FCC
Statement
EMC
Guidance
24