Intelligent Measuring System

ali-Q 2’s 

aliquot mode 

utilizes a state-of-the art measuring system that incorporates information and feedback from 

multiple sensors to give real-time feedback and parameter adjustment for different variables that affect the accuracy 

and precision when in 

aliquot mode

-

Varying environmental conditions

: Temperature, humidity and pressure all have significant effects on the

performance of any liquid handling device…. EXCEPT for the ali-Q 2! ali-Q 2 can sense changes in ambient conditions 

and compensate as necessary to ensure accurate aliquoting whether you're in the deserts of Arizona, or the 

mountains of Colorado! 

-

Varying pipet angle

: One of the most useful (and ergonomic!) features of the ali-Q 2 is its ability to adjust the

dispensed aliquots for varying pipet angles. This is especially useful in a hood, where space and maneuverability are

limited. Whether you hold the pipette completely vertical or tilted, you can be confident that accuracy and

repeatability of the aliquots will not be affected.

-

Varying brands/sizes serological pipets: 

Every brand of pipet has different sizes and geometries. These geometric

differences contribute to changes in the physical characteristics of the system. ali-Q 2 can sense the effect of these

differences and will automatically compensate accordingly to optimize performance.

Wall Mount:  

ali-Q 2 may be stored in the wall mount (included in box) to avoid damage from accidental drops and chemical spills 
(and save precious bench space!).  ali-Q 2 may also be placed in the wall mount with pipets up to 100 mL inserted. The 
wall mount will firmly adhere to any 

smooth, clean, non-painted 

surface, such as the inside of a hood, using the 3M 

Command strips included. Clean the surface with alcohol before applying the wall mount. Remove the stickers on the 
adhesive backs and immediately press onto surface and keep applying moderate pressure for about 30 seconds. DO 
NOT place ali-Q 2 into the wall mount for about 30 minutes to allow the adhesive to set. 

Kickstand:  

ali-Q 2 can be used with the removable kickstand (included in box) even with a serological pipet inserted. The 

kickstand attaches securely in one orientation only. Be sure to attach so that the curved ends point up. Simply snap 

the pins on the kickstand into the round dimples on the sides of the ali-Q 2 body. To remove, gently pinch the curved 

ends and pull kickstand away from body. 

Kickstand 

Содержание 2100-0003

Страница 1: ...ali Q 2 Series Operating Manual...

Страница 2: ...tem 9 Wall Mount 9 Kickstand 9 Technical Specifications 10 Calibration 11 In lab Calibration Protocol 12 Initial PerformanceVerificationWorksheet 12 Troubleshooting 14 Maintenance 15 Cleaning Decontam...

Страница 3: ...8 Use only VistaLab accessories and spare parts See Ordering Information on pg 18 Do not attempt to make any technical alterations Do not dismantle the instrument any further than is described in the...

Страница 4: ...arging port USB C N A 4 Aliquot button Nylon 5 Aspirate button Nylon 6 Dispense button Nylon 7 Battery Lithium ion Lithium Polymer 8 Battery compartment cover PC ABS 9 Nozzle Cover Nylon 10 Nozzle Ins...

Страница 5: ...ional part of a total amount of a solution or suspension verb To divide as a solution into equal parts Aspirate verb to draw by suction Dispense verb to expel liquid Serological Pipet pipet noun dispo...

Страница 6: ...Mode see next page the speed that the ali Q 2 aspirates and dispenses liquid is controlled by the touch sensitive aspirate and dispense buttons The amount of force used to push on the buttons determin...

Страница 7: ...the bottom gray dispense button until bottom of meniscus is at the desired volume graduation ATTENTION Do not fill the serological pipet over the max volume Severe damage can occur if any liquid ente...

Страница 8: ...as you will need Ex If you need to dispense 10 aliquots of 2 mL you will need to aspirate at least 20 mL of liquid 5 Aliquot Carefully position the tip of the full pipet over the first receiving vess...

Страница 9: ...has different sizes and geometries These geometric differences contribute to changes in the physical characteristics of the system ali Q 2 can sense the effect of these differences and will automatica...

Страница 10: ...racy Precision CV 3 0 mL 2 1 0 3 mL 3 3 With any other brand of serological pipet any size Dimensions 7 2 x 7 0 x 2 2 18 3 cm x 17 8 cm x 5 6 cm Weight 350g Volume Capacity Accepts 1 to 100 mL serolog...

Страница 11: ...ends sending ali Q 2 to VistaLab for calibration or performing verification at least once a year or whenever conditions and equipment change In lab Verification Adjustment Users can verify the accurac...

Страница 12: ...e initial current error at 0 5 and 5 0 mL and record the 2 numbers to use in the following calibration procedure Worksheet Determine the current accuracy of your ali Q 2 by taking the average of 4 gra...

Страница 13: ...dispensing too high 1 5 error press the Aspirate button to save the changes If the ali Q 2 is dispensing too low 1 5 error press the Dispense button to save the changes The 2 LEDs will alternate blue...

Страница 14: ...filter properly see pg 4 Replace nozzle insert Send ali Q 2 in for repair service Pipet does not stay in ali Q 2 Nozzle insert worn or damaged Replace nozzle insert NOTE Using Wobble not pipets will...

Страница 15: ...54 nm wavelength from UV sterilization lamps NOTE The effects of the UV exposure will likely cause discoloration over time Autoclaving NEVER AUTOCLAVE ALI Q 2 Only the nozzle cover nozzle insert and m...

Страница 16: ...the power supply pack and insert into wall outlet 2 Attach the wide end of the USB C charging cord to the power supply pack 3 Plug the USB C end of the charging cord into the charging port of the ali...

Страница 17: ...hnical Support to receive Return Authorization Unit will be inspected for defects and repaired replaced as necessary at the sole discretion of VistaLab Customer will be responsible for service charges...

Страница 18: ...terile 5 pack 9020 4006 Replacement Membrane Filter 0 45 m sterile 5 pack 9020 4008 Replacement Silicone Nozzle Insert 2 pack 9020 4010 Replacement Nozzle Cover 1 pack 9020 4011 Replacement Kickstand...

Страница 19: ...of their service life According to the Directive 2002 96 EC of the European parliament and of the Council on Waste Electrical Equipment WEEE of 27 January 2003 electronic equipment requires disposal a...

Страница 20: ...rical equipment for measurement control and laboratory use General Requirements CAN CSA C22 2 No 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use General...

Отзывы: