Troubleshooting
If problem still occurs, contact VistaLab Technical Support at (914) 244- 4068 or [email protected]
Problem
Possible Cause
Suggested Fix
Slow aspiration
-
Membrane filter clogged
-
Not enough force on aspirate button
-
Replace membrane filter
-
Press aspirate button with more force
to increase aspirate speed
Aspirating air bubbles
-
Pipet tip not fully immersed in liquid
while
aspirating
-
Pipet not fully inserted into nozzle insert
-
Aspirating too fast for solution
(ex: detergents)
-
Make sure pipet tip is fully immersed in
liquid throughout aspiration
-
Insert pipet more
-
Lightly press the aspirate button to slow
aspirate speed
Pipet
te
drips liquid
-
Pipet not fully inserted into nozzle insert
-
Crack in serological pipet
-
Membrane filter not properly
positioned
or upside down
-
Nozzle insert worn or damaged
-
Valve not shutting properly
-
Try Wobble-not
low insertion force,
less drip serological
pipets
-
Get a new pipet
-
Insert filter properly
(see pg. 4)
-
Replace nozzle insert
-
Send ali-Q 2 in for repair service
Pipet does not stay in ali-Q 2
-
Nozzle insert worn or damaged
-
Replace nozzle insert
NOTE: Using Wobble-not pipets will
significantly
decrease
damage to the
nozzle insert.
Out-of-Calibration
-
Using different equipment
or
different
density liquid, varying environmental
conditions
-
Membrane filter clogged
-
See In-lab Calibration on page 12
-
Replace membrane filter
ali-Q 2 makes “error” sound
when dispensing aliquots
-
Not
holding
down aliquot button long
enough
-
HOLD aliquot button until
status
LED
blinks
GREEN
and motor starts (you
will hear and feel motor)
LED blinks or does not light up
when connected to charger
-
Battery not properly installed
-
Wrong Power supply
-
Battery / Power supply damaged
-
Firmly push in battery connector
-
Only use Power supply Cat. #9020-9005
-
Replace battery/ replace Power supply
Battery does not hold charge
-
Charger not inserted fully
-
Battery defective or drained
-
Check charger connection. Replace
charging cord
-
Replace battery
ali-Q 2 is unresponsive
-
Dead battery
-
Recharge battery
-
Restart unit by unplugging battery for 10
seconds and plug battery back in.