GB-47
english
Appendix
G
Malfunc-
tions
Possible reasons
Corrective action
Err -1
Movement of the arm or body during the meas-
urement. Do not talk during the measurement.
Repeat the measurement after 3-5 minutes. Don’t move
the arm and don’t talk. In case of recurrence please
contact our after-sales service.
Err -2
Pump problem
Check whether the air connection is correctly inserted. Is
the cuff
fi
tted too loosely? Repeat measurement. In case
of recurrence please contact our after-sales service.
Err - 3
De
fl
ation rate too high or too low, valve problem
In case of recurrence please contact our after-sales service.
Err - 300
Upper pressure limit of 300 mmHg has been
exceeded.
Repeat the measurement after 3-5 minutes rest
Irregular heartbeat, disruptive movements,
shaking, wobbling (objects), breathing in deeply,
etc.
Repeat measurement after 3-5 minutes rest. Possible
factors (depending on the severity of the arrhythmia)
affecting the measurement results should be discussed
with your doctor.
Levels too
high
Was the necessary rest observed before
measurement?
Repeat measurement after a break of approx. 3-5 min.
Do not move your arm and do not talk.
Unusual
levels
Movement or talking during measurement. Rest-
ing time not observed, feet possibly crossed,
refrain from smoking or drinking coffee at least
30 minutes before measurement.
Check conditions and repeat measurement.
Содержание comfort III
Страница 6: ...deutsch D 6 Gerätebeschreibung A ...
Страница 27: ...D 27 deutsch Anhang G ...
Страница 32: ...english GB 32 Device description A ...
Страница 53: ...GB 53 english Appendix G ...
Страница 58: ...français F 58 Description de l appareil A ...
Страница 79: ...F 79 français Annexe G ...
Страница 84: ...italiano I 84 Descrizione dello strumento A ...
Страница 107: ...I 107 italiano Appendice G ...