Retrofitting the Z axis module

  This procedure should only be carried out by  

authorised service personnel. 

 

Please contact your local Vision Engineering 

branch for assistance

Optional riser block

  This procedure should only be carried out by  

authorised service personnel. 

 

Please contact your local Vision Engineering 

branch for assistance

Episcopic LED replacement

 

`

Remove the episcopic assembly 

u

 from the Hawk  

and carefully place it on a workbench as shown.

 

`

Pull its power lead 

v

 clear of the socket.

 

`

Loosen the 2 securing screws 

w

 and pull the LED  

assembly 

x

 clear of the illuminator.

 

`

The replacement procedure is the reversal of the  
above.

u

v

w

v

u

w

x

Содержание Swift PRO Cam

Страница 1: ...ft PRO Cam 9 9 Swift PRO Elite User Guide Vision Engineering Ltd has been certified for the quality management system ISO 9001 2015 FM 557119 Est 1958 www visioneng com Swift PRO series of optical and...

Страница 2: ......

Страница 3: ...safety A dust cover is supplied with your system Turn off your system and ensure it has cooled down before using the dust cover Illumination safety Do not look directly at the illuminated LED s This...

Страница 4: ...ision Engineering Ltd The Freeman Building Galileo Drive Send Woking Surrey GU23 7ER UK www visioneng com Disclaimer This document is provided as is without any warranty of any kind either express or...

Страница 5: ...C system connection 6 QC 3000 connection 6 Stage alignment video systems only 7 Securing the stage 150mm x 100mm 200mm x 100mm 7 Head attachment 8 Fitting the stage glass 8 Stage glass levelling 9 Ant...

Страница 6: ...Height adjustment 14 Encoder Resolution 14 Video Camera 14 Optics Swift PRO Duo and Elite 14 Illumination 14 PSU fuse ratings 14 Dimensions 14 Swift PRO Duo and Elite 14 Swift PRO Cam 14 Service calib...

Страница 7: ...com support 1 System equipment Stand objectives Stand Lens mount Lens options Illuminator options Toolkit Anti glare shield Z Axis upgrade kit optional Swift PRO Cam Stage 150mm x 100mm manual stage...

Страница 8: ...System equipment www visioneng com support 2 Head QC 3000 Controller PC Monitor Keyboard and mouse...

Страница 9: ...of your system It is highly recommended that you refit the transit protection whenever you transport your system Attaching the lens mount Insert the lens mount u up into the focus assembly v and secu...

Страница 10: ...g the episcopic illuminator EPI Locate the cone u over the EPI v and secure it with the grub screw w Locate the EPI into the focus assembly x and secure it with the grub screw y Connecting the EPI Ins...

Страница 11: ...the underside of the stage x are clean and free of any debris Loosely fit the stage bolts y in position B1 B2 and B3 also in diagram v to a light finger tight tension If this is a video system do not...

Страница 12: ...gaged fully and secured WARNING To comply with safety regulations easy access to the mains socket must be maintained Y Axis Z Axis if fitted From camera USB to controller USB to controller X Axis Main...

Страница 13: ...n be removed the relevant stage securing procedure should be carried out Securing the stage 150mm x 100mm 200mm x 100mm With the stage assembled see page 5 and aligned see above secure the stage as fo...

Страница 14: ...rn the head to align the head crosshair with the alignment plate see inset Tighten the securing screw Remove the alignment plate If you need to remove the stage for any reason re attach the alignment...

Страница 15: ...the front right hand corner into view Use the relevant adjustable glass support to bring the surface of the glass into sharp focus Repeat for the remaining 2 corners Repeat the above steps if necessar...

Страница 16: ...Y axis control X axis control Coarse focus control Fine focus control On off switch Substage light dimmer Episcopic light dimmer Ringlight dimmer Iris control All desktop control systems configured b...

Страница 17: ...pply Maximum relative humidity of 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C This system is an accurate industrial gauging instrument To achieve the optimum acc...

Страница 18: ...e the graticules v 2 off into position and secure them with the lock nuts and grubscrews w 2 per graticule and the retaining bolts x To focus and centralize each graticule loosen the appropriate retai...

Страница 19: ...cedure should only be carried out by authorised service personnel Please contact your local Vision Engineering branch for assistance Episcopic LED replacement Remove the episcopic assembly u from the...

Страница 20: ...ideo camera Optics Swift PRO Duo and Elite Patented twin pupil monoscopic infinity corrected optical system with pre centred crossline graticule to both eyes Option of custom designed graticule pre ce...

Страница 21: ...ervice calibration record Serial numbers System serial number Stage serial number PSU serial number Service information Service type Comments Date of service Date of next service Company Signature Sys...

Страница 22: ......

Страница 23: ...in transit damage caused by misuse neglect or carelessness or damage resulting from either improper servicing or modification by other than Vision Engineering approved service personnel Further this...

Страница 24: ...sioneng com Vision Engineering Ltd Italia Via G Paisiello 106 20092 Cinisello Balsamo MI Italia Tel 39 02 6129 3518 Email info visioneng it Vision Engineering South East Asia P 03A 20 Impian Meridian...

Отзывы: