Visiomed – EasyScan – User Manual V.4-102008
24/39
T
ROUBLESHOOTING
Troubles
Checklists
Countermeasures
Battery used up?
Change new battery
No response
Battery in wrong polarity?
Adjust battery to proper
polarity.
Battery symbol on LCD
Low battery
Change new battery
“Lo” on screen
(result below 32°C or
89.6°F)
Thermometer rightly
pointed at the eardrum /
forehead?
Follow User’s Manual to
point at the eardrum /
forehead.
“Hi” on screen
(above 42.9°C or
109.3°F)
Please check the operation
method
Follow User’s Manual for
proper measurement.
“ErrE” on screen
Ambient temperature out
of range
Do not use the
thermometer in extreme
temperatures
“ErrU” on screen
Cannot measure
Follow steps in User’s
Manual for proper use.
“ErrH” on screen
Hardware problem
Consult your dealer.
“ErrS” on screen
Sensor error
Do not use the
thermometer in extreme
temperatures
“ErrR” on screen
Hardware problem
Consult your dealer.
“ErrL” on screen
Battery low.
Replace battery.
“ErrP” on screen
Hardware problem
Consult your dealer.
S
PECIFICATIONS
Product type
Medical infrared talking thermometer
Model
EasyScan FTX3
Temperature sensing part
IRsensor
Using Condition
15°C to 40°C
Storage condition
Temperature: From -10°C to 55°C
Humidity rate: 20 % HR~85%HR
Battery
1 button cell batteryCR2032
Dimension
109 × 44,6 × 28,4 mm
Weight
80 g (without battery)
Temperature display
°C / °F
Body range
From 32,0°C to 42,9°C
Room range
From 5,0°C to 59,9°C
Precision
+/-0,2°C
Power consumption
< 50 mA (for voice prompt)
Automatic stop