BewellConnect - BW-GL1 - User Manual - 052015
33
• practicing the testing process, or
• you have dropped or think you may have damaged the meter.
Test strips
(c)
, control solutions
(d)
, lancing device
(e)
or sterile lancets
(f)
may not be includ-
ed in the kit (please check the contents on your product box). They can be purchased sepa-
rately. Please make sure you have those items needed for a blood glucose test beforehand.
5.2. PERFORMING A CONTROL SOLUTION TEST
To perform a control solution test, you will need:
(b)
,
(c)
and
(d)
.
1. INSERT THE TEST STRIP TO TURN ON THE METER
Insert the test strip into the meter. Wait for the meter to display the “
” and “ “.
2. PRESS M TO MARK THIS TEST AS A CONTROL SOLUTION TEST
With “CTL” displayed, the meter will not store your test result in memory under “CTL”. If
you press
M
again, the “CTL” will disappear and this test is no longer a control solution
test.
WARNING:
When doing the control solution test, you have to mark it so that the test result will NOT
be stored in the memory. Failure to do so will mix up the blood glucose test results with the
control solution test results in memory.
3. APPLY CONTROL SOLUTION (G)
Shake the control solution vial thoroughly before use. Squeeze out the first drop and
wipe it off, then squeeze out another drop and place it on the tip of the vial cap. Hold
the meter to move the absorbent hole of the test strip to touch the drop. Once the
confirmation window fills completely, the meter will begin counting down.
NOTE: To avoid contaminating the control solution, do not directly apply control solu-
tion onto a strip.
4. READ AND COMPARE THE RESULT
After counting down to 0, the control solution test result will appear on the display.
Compare this result with the range printed on the test strip vial and it should fall within
this range. If not, please read the instructions again and repeat the control solution test
NOTE:
• The control solution range printed on the test strip vial is for control solution use only.
It is not a recommended range for your blood glucose level.
• See the MAINTENANCE section for important information about your control solutions.
Содержание BewellConnect MyGluco BW-GL1
Страница 2: ...BewellConnect BW GL1 User Manual 052015 1 BW GL1 MyGluco 0123...
Страница 4: ...BW GL1 MyGluco 0123 FR...
Страница 6: ...BewellConnect BW GL1 User Manual 052015 7 a g c d e f b h i 1 2 m l j k...
Страница 24: ...BW GL1 MyGluco 0123 EN...
Страница 26: ...BewellConnect BW GL1 User Manual 052015 27 a g c d e f b h i 1 2 m l j k...
Страница 42: ...BW GL1 MyGluco 0123 ES...
Страница 44: ...BewellConnect BW GL1 User Manual 052015 47 a g c d e f b h i 1 2 m l j k...
Страница 62: ...BW GL1 MyGluco 0123 NL...
Страница 64: ...BewellConnect BW GL1 User Manual 052015 67 a g c d e f b h i 1 2 m l j k...
Страница 81: ...BW GL1 MyGluco 0123 DE...
Страница 83: ...BewellConnect BW GL1 User Manual 052015 87 a g c d e f b h i 1 2 m l j k...
Страница 100: ...BewellConnect BW GL1 User Manual 052015 104 WA R R A N T Y...