
32
Is mounted in the center of the detector end plate. Below the source clamp, the detector block
is mounted to the detector end plate,
Figure 11
. Inside the detector block is the beam splitter
prism and the two photodiode detectors, located on either side of the detector block. The
photodiode detectors are also mounted via their leads to the photodetector circuit board (PCB).
Cable connections are provided on the photodetector PCB for the LED light source and
for the I/O cable to the RI Output PCB. Enclosing the source/detector end is the blackened
aluminum endcap. A slit is provided in the cap for the I/O cable to exit.
RI OUTPUT
Outputs from the two photodetectors are differentially acquired on the TIA photodetector
PCA. This differential signal is fed via the I/O cable to the main CPU board
RI output voltage is displayed on the front panel display meter when the detector is
selected from the front panel board and is also available as DC output voltage on the back
panel output connector when selected. The RI photoelectric signals are not adjustable by the
user.
OPTICAL ZERO
The user can adjust this optical zero approximately by using the front panel switch to
adjust the zero glass assembly. Please see section A.1.b General Layout for operation.
REFERENCE PURGE
The reference cell can be purged under normal flow conditions by switching ON the RI
purge valve. Continue solvent flow for several minutes or until a stable RI signal is obtained.
The flow path during purge condition can be followed from the flow diagram. Flow is
always first through the sample cell, then through a Tee
to the top of the reference cell, through
the purge valve, then out to the viscometer.
VAPOR SENSOR
The vapor sensor can be adjusted more or less sensitive depending on the solvent used.
A level detector circuit (U21) monitors the chemical sensors output signal. The detection level is
determined by potentiometer VR5 on the 301 CPU board. The factory setting for VR5 is 4.0
VDC. Clockwise rotation of VR5 increases the sensitivity of the detector. The voltage setting of
VR5 can be measured across C1 on the CPU board, which is located directly below VR5. It
should be noted that the chemical sensor has a built in heater, which requires a warm-up period
of approximately 5 minutes. During the initial power-up state, the vapor detector output signal
will be higher than its operating steady state level. Should a false alarm be detected, turn VR5
counter-clockwise to decrease the sensitivity of the detector.