Calibrating
A calibration adapter is available for calibrating the eco-DUO 450. It is mounted on the
outlets. The eco-Control EC 200-DUO calibration programmes can be used to determine
the exact dosing quantities and save them for calculations in the dosing programmes.
9
35
9
11
Calibrate the dispenser
• Remove the static mixing pipe.
• Unscrew the screws (9) in the locking plate
(11).
• Mount the calibration adapter (35)
• Carry out the calibration in accordance with the
eco-Control EC 200-DUO operating manual.
Once calibration is complete, dismantle the
calibration adapter and position the static mixing
pipe.