eco-CONTROL EC200-B
Operation and maintenance instructions
The set value remains active until the next change or until the Dosing systems is
switched off. Alternatively, a value of its own for the flow rate can be assigned to each
dosing program (See 5.14, Changing a dosing program, page 29).
5.10.4 Sucking back
In order to effectively prevent any dripping of the medium, a value for the
sucking back can be set. The maximum amount for sucking back that can
be set is 0.5 ml (eco-PEN300 50 µl).
• Selecting, changing and saving the displayed value.
The sucking back is done after each dosing operation.
D
is
pl
ay
3
5
The set value remains active until the next change or until the Dosing systems is
switched off. Alternatively, a value of its own for the sucking back can be assigned to
each dosing program (See 5.14, Changing a dosing program, page 29).
5.10.5 Material density
Inputting the material density for the first time
• Press the Quantity program shortcut key, menu Quantity
program, display 3 appears.
• Confirm selection, display 4, input dialog for material
density appears.
Note:
If the quantity can be switched from
ml
(
µl
)
to
g
(mg)
, a
material density has already been set and saved. See the
following description
Changing the set material density.
D
is
pl
ay
3
• Set, confirm and save the material density.
After leaving the input dialog, the
Quantity program
menu,
display 3, appears again.
D
is
pl
ay
4
Quantity program
Esc
Suckback
Flow rate
Volume
0%
100%
ml/min
00.100
03.000
00.000
ml
ml
Esc
0.000
Density of material
Insert density
Density
OK
g/cm³
kg/dm³