FRmt: Magnitude Frequency Response (MultiTones, 32 1/3 Octave Bands)
A multitone signal consists of 32 single frequency tones corresponds to the center
frequencies of 32 1/3 octave bands will be generated by pressing the start button of the
signal generator. The signal will be injected into the DUT (Device Under Test), and the
response of the DUT will be captured by the oscilloscope and spectrum analyzer. The
curve in the spectrum analyzer indicates the magnitude frequency response of the DUT.
(19)FRph: Magnitude Frequency Response (Frequency Sweep, Linear, Peak Hold)
A 60-second 10Hz-to-25kHz linear frequency swept sine wave will be generated
repeatedly by pressing the start button of the signal generator. The signal will be injected
into the DUT (Device Under Test), and the response of the DUT will be captured by the
oscilloscope and spectrum analyzer. The curve in the spectrum analyzer indicates the
magnitude frequency response of the DUT.
(20)FR100k: Magnitude Frequency Response (0~100kHz, White Noise, DDS Mode, Avg. 30)
White noise will be generated by pressing the start button of the signal generator. The
signal will be injected into the DUT (Device Under Test), and the response of the DUT
will be captured by the oscilloscope and spectrum analyzer. The curve in the spectrum
analyzer indicates the magnitude frequency response of the DUT. Unlike the settings
from (2) to (19), which deal mainly with audio frequency, the setting here can be used to
measure frequency response from 0 to 100kHz.
www.virtins.com 29 Copyright © 2017 Virtins Technology