+Taking Measurement
Operation Procedure
There are three different methods of taking the measurement. Start with
measurement method A, “right index finger-chest.” If this method does not
deliver measurements, or delivers only unstable measurements (“EE” is
frequently displayed), switch to method B “left index finger-chest” and, if
necessary, to method C ”left hand-right hand.”
The most suitable method/ procedure depends on the heart configuration
(shape of the heart) of each individual user. If it is not possible to take stable
measurements using a given measurement method, this could have a
harmless cause such as the shape of the heart. However, the cause may also
be an illness or disease.
Note:
Measurement method C offers maximum comfort, but offers a much
lower measurement stability than methods A or B.
Press the on/ off button for about three seconds to turn the device on. The
device shows the full display.
WARNING
If segments are missing, stop using the device and immediately
contact customer services. To test whether the full display is
completely displayed, hold the On/Off button when switching on the
device.
+Taking Measurement
Measurement Method A
“right index finger-chest” (approximately corresponds to “lead 2”)
1. Press the power button for 3 seconds to turn on your ECG Pen.
2. Lay your right index finger on the two top electrodes of the device
and hold the ECG Pen face up in your hand.
3.
You can determine the correct position for placing
the lower electrode of the device against your chest
using the following methods:
Place the bottom electrode of the ECG Pen on
bare skin about 5 centimeters (2 inches) below your left nipple supported
by the ribcage, and press it
GENTLY
until you hear the click sound.
Draw an imaginary line from the front of your armpit downwards. At
the same time, draw an imaginary line going up 10cm from the lowest rib on
the left side of your body. Place the lower electrode of the device at the
point where these two lines meet.
Draw an imaginary line from the bottom of the centre of your chest
(sternum) to the left. At the same time, draw an imaginary line from the
front of your armpit downwards. Place the lower electrode of the device at
the point where these two lines meet.
Содержание 870E
Страница 24: ...Vion ECG Pen 880E 870E ...
Страница 29: ...MD Biomedical Inc 8F No 222 Sec 4 Chengde Rd Taipei City 111 Taiwan www mdbiomedical com 35 ...
Страница 30: ......