8
General Characteristics (cont.)
D122 T132
Name Description
*
Disabled Indicator
Red LED indicates when interrupt is activated. Shutter
driver is disabled. Interrupt out is Active-high.
*
*
Pulse Energy Select
Allows selection of additional pulse energy necessary when
using the VS35 shutter. Selected by HIGH/LOW toggle
switch located on the rear panel.
*
*
Power Requirements
110-120/220-240VAC 60/50 Hz. Selected automatically by
internal control circuit.
*
*
Fuse Requirements
3AG ½ amp SB for 115VAC line. 3AG ½ amp SB for
shutter protection. For 230VAC operation change line fuse
to ¼ amp SB. (SB indicates slow blow or time delay fuse.)
*
*
Size (HWD)
2.64” (67.0 mm) x 5.35” (136.0 mm) x 8.19” (207.9 mm)
with black textured finish. See figure #7.
Accessories
D122 T132
Name Description
* *
Accessories
(supplied)
Model #
710C
Shutter Interconnect Cable 10 ft length. 7-pin
male connector to 7-pin female connector.
Accessories
(optional)
Model #
710R
Remote Trigger Cable 10 ft length. Hand
held push button switch to BNC connector.
Accessories
(optional)
Model #
710R/F
Remote Trigger Cable 10 ft length. Foot
activated switch to BNC Connector.
All BNC inputs can be activated by connecting ground to the specified BNC input. (The 710R cable will
accomplish this. Connect the BNC of the 710R cable to any input and that control function will be activated when
the pushbutton is depressed.)
*CAUTION – DO NOT connect the 710R cable to an output, this may cause irreparable damage to the
specified output.
All control inputs (terminal strips) can be activated by switching appropriate DC voltages (+5VDC or +8VDC) to
the specified input.
Any means of switching (electronic or mechanical) may be used to activate the inputs described above, provided
all input specifications for a particular input are noted.