USER'S MANUAL
Operating instructions
ROTOGRAPH-D (120V)
(Rev. 1)
40
6.5
6.5
6.5
6.5
Messages on display
Messages on display
Messages on display
Messages on display (refer to Figure 7 and
Figure 8)
ROTOGRAPH-D is entirely controlled by a microprocessor that, besides
controlling the emission parameters and the line voltage fluctuation,
indicates the different equipment's operating modes by means of
appropriated messages on display (44), the eventual anomalies and
mistakes. The following table shows the different messages that could be
displayed, their cause and relevant actions.
Message displayed
Reference
Cause
What to do
Minimum
0
With control button
(40) it has been
required a value lower
than the allowed one
Select a correct value
Maximum
0
With control button
(41) it has been
required a value higher
than the allowed one
Select a correct value
Wait, please.....
6.2.4,
6.3.1.4 and
6.3.2.3
X-ray exposure has
been performed
Waiting time (4min) for
PAN or TMJ exposures
Equipment not set
6.2.1 and
6.3.1.1
Not all interlocks are
enabled
Press a second time the
function button to
display the interlock to
be enabled
Digital Sensor Not
ready
6.2.1. step 1 Digital sensor not
enabled
Request activation of
digital sensor by the
QuickVision program
Select PAN/TMJ1
NO reference Functionality of the
digital sensor has not
been selected
Select functionality of
the sensor activating
one of the buttons on
the cover of the digital
sensor
Select TMJ2
NO reference The sensor is not
activated for TMJ2 after
having selected this
test on the hand
control
Select functionality
TMJ2 of the sensor
activating the button
(15) on the cover of the
digital sensor
Line voltage too low
NO reference Line voltage outside the
-10% of the nominal
value
(1)
Make sure the line
voltage is within the
±10% of the nominal
value
Line voltage too high
NO reference Line voltage outside the
+10% of the nominal
value
(1)
Make sure the line
voltage is within the
±10% of the nominal
value
Содержание ROTOGRAPH-D
Страница 3: ...USER S MANUAL Revision history ROTOGRAPH D 120V Rev 1 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 36: ...USER S MANUAL Operating instructions Rev 1 ROTOGRAPH D 120V 31 9 18 20 24 45 46 7 Figure 4 TMJ1 position ...
Страница 40: ...USER S MANUAL Operating instructions Rev 1 ROTOGRAPH D 120V 35 9 18 20 24 45 46 7 Figure 5 TMJ2 position ...
Страница 49: ...USER S MANUAL Operating instructions ROTOGRAPH D 120V Rev 1 44 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 55: ...USER S MANUAL ROTOGRAPH D 120V Rev 1 52 THIS PAGE IS INTENTIONALLY LEFT BLANK ...