![videomed Truelink 4 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/videomed/truelink-4/truelink-4_instructions-for-use-manual_1015572018.webp)
Safety information
18
80028008_030_B – 773625 – 2021-04-29
CAUTION
MEASUREMENT OF DISPERSAL CURRENTS!
It is necessary to measure the leakage currents with the circuits
downstream of the Truelink 4 system open. Otherwise, the
leakage currents of these circuits will be added to those of the
Truelink 4 system.
3.2
Electromagnetic compatibility
The Truelink 4 Video Management System supplied contains
electronic components subject to the Electromagnetic
Compatibility regulations, affected by conducted and radiated
emissions.
The emission values comply with regulatory requirements thanks
to the use of components compliant with the Electromagnetic
Compatibility Directive, suitable connections and installation of
filters where required.
The Truelink 4 Video Management System is therefore compliant
with the Electromagnetic Compatibility (EMC) directive.
CAUTION
Any maintenance activities on the electrical equipment carried
out in a non-compliant manner or incorrect replacement of
components may compromise the efficiency of the solutions
adopted.
The Truelink 4 product is a Class A electromedical device
according to CEI 60601-1-2 (CISPR 11), it is suitable for use in a
specific electromagnetic environment. The customer and / or the
user of the product must ensure that it is used in an
electromagnetic environment as described below.
Emission Test
Compliance
Electromagnetic Environment Guide
Radiated and conducted RF
emission
CISPR 11
Group 1
Truelink 4 uses RF (radio-frequency) energy
only for its internal operation. Therefore, the
RF emissions are very low and should not
cause interference in adjacent electronic
devices.
Class A
Truelink 4 is suitable for use in all buildings,
except domestic buildings and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Not applicable