Guidance and manufacturer’s declaration – electromagnetic immunity
Table 204
The DiagnosticPRO
Edge
Digitizer is intended for use in the electromagnetic environment specified below. The
customer or end user of the DiagnosticPRO
Edge
Digitizer should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level Electromagnetic environment - guidance
Portable and mobile RF communication equipment should
be used no closer to any part of the DiagnosticPRO
Edge
Digitizer, including cables, than the recommended
separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
P
d
2
.
1
80 MHz to 800 MHz
P
d
3
.
2
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked
with the following symbol:
Note 1: At 80MHz and 800MHz, the higher frequency range applies.
Note 2: These guidelines may not be applicable in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the
DiagnosticPRO
Edge
Digitizer is used exceeds the applicable RF compliance level above, the
DiagnosticPRO
Edge
Digitizer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the Digitizer.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
d
2
.
1