VICI VC837 Скачать руководство пользователя страница 2

6V

±(0.8%+6)

1mV

60V

10mV

600V

±(1.0%+6)

100mV

Input impedance: 10MΩ.

Overload protection: 600V DC/ AC rms.

Frequency response: at 600V range: (40~1000)Hz, at other ranges: (40~2000)Hz

Displaying: True RMS response (calibration based on sine wave RMS)

3.2.4 DC Current

DCA

Range

Accuracy

Resolution

600uA

±(1.0%+5)

0.1μA

6000uA

1μA

60mA

10μA

600mA

100μA

6A

1mA

10A

±(2.0%+5)

10mA

Maximum voltage drop: 200 mV for full range.

Maximum input current: 10A (within 10 seconds).

Overload protection: 0.5A/250V fuse and 10A/250V fuse.

3.2.5 AC Current (ACA)

Range

Accuracy

Resolution

600uA

±(1.5%+5)

0.1μA

6000uA

1μA

60mA

10μA

600mA

100μA

6A

1mA

10A

±(2.0%+10)

10mA

Maximum voltage drop: 200 mV for full range.

Maximum input current: 10A (within 10 seconds).

Overload protection: 0.5A/250V fuse and 10A/250V fuse.

Frequency response: 10A range: 40~100Hz, other ranges: 40

400Hz.

3.2.6 Resistance

Ω

Range

Accuracy

Resolution

600Ω

±(0.8%+5)

0.1Ω

6kΩ

±(0.8%+1)

60kΩ

10Ω

600kΩ

100Ω

6MΩ

1kΩ

60MΩ

±(1.2%+5)

10kΩ

Open circuit voltage: 500mV

Overload protection: 250V DC/AC peak value.

3.2.7 Capacitance (C)

Range

Accuracy

Resolution

60nF

±(2.5%+6)

10pF

600nF

±(2.5%+5)

100pF

Содержание VC837

Страница 1: ...n and range to avoid wrong operation 2 5 Move the test leads away from test points when switching to other function 2 6 Don t input voltage in current terminal 2 7 Don t make any modification to the c...

Страница 2: ...onds Overload protection 0 5A 250V fuse and 10A 250V fuse 3 2 5 AC Current ACA Range Accuracy Resolution 600uA 1 5 5 0 1 A 6000uA 1 A 60mA 10 A 600mA 100 A 6A 1mA 10A 2 0 10 10mA Maximum voltage drop...

Страница 3: ...1832 750 1 0 5 750 1 5 15 Thermocouple K type WARNING do not input voltage at this range Operation 4 1 Panel Description LCD Display the data and unit symbol Number Feature Description 1 Relative REL...

Страница 4: ...e is no measurement in 15 minutes the meter will automatically power off and enter sleep mode One minute before sleep mode the buzzer will beep for 5 times to remind user Press any button or turn the...

Страница 5: ...s the default when you turn on the meter AUTO symbol displayed Press RANGE REL key can change to Manual range the available ranges are 600uA 6000uA 60mA 600mA 6A 10A 4 4 3 Connect test leads to the te...

Страница 6: ...ance value NOTE 1 Don t input voltage or current under capacitance ranges 2 Press RANGE REL to clear the reading before testing to assure the accuracy 3 There is only the auto range mode under the cap...

Страница 7: ...RATURE MEASUREMENT 4 11 1 Turn the rotary switch to range Press SELECT key to select or mode 4 11 2 Insert the cold end free end of thermocouple in V mA and COM terminal and put the working end temper...

Страница 8: ...ter does not work properly please check the meter as following steps If the problems still cannot be solved please refer to repairing center or contact the local dealers Fatult Solution No reading on...

Отзывы: