7-1 Error Messages
-75-
The balance is influenced by static
electricity or noise.
- Unplug the AC adapter from the
receptacle and then turn the power
on again.
- If this error occurs again, electric
components may have failed.
Contact our Sales Office or
Technical Service Division.
The weight of a sample is too light at a
sampling during parts counting, or reference
weight saving during percentage weighing.
Use a heavier sample by referring to
the Specifications (P. 88) to check the
minimum unit weight and the
percentage weighing weight limit.
- At addition operation, you placed
additional samples on twice.
- At addition operation, you unloaded some
samples or you pressed the key without
adding samples.
- After setting the display to “0” (by
unloading the previous sample),
place the next sample to continue
addition operation.
- Addition operation is impossible
when 0 or a negative value is
displayed. Place a sample to
continue addition operation.
No inputs are sent from the weight sensor.
- Unplug the AC adapter from the
receptacle and then turn the power
on again.
- If this error occurs again, the sensor
may have failed. Contact our Sales
Office or Technical Service Division.
Because the balance is unstable,
initialization cannot be completed.
The balance may be affected by an
external influence such as wind and
vibration. Relocate the balance by
referring to the section “Do not use the
balance in the following places” (P. v).
Cannot initialize because the power was
turned on with an object on the pan.
Please remove the object from the pan.