8.  Limit Function 

The limit function judges measurements according to a limit value saved in the balance.  

The function shows the judgement result by displaying the [

] mark as either HI (excessive), OK 

(appropriate), or  LO (insufficient).  This function is very useful when discriminating between conforming 
and nonconforming articles.  It is also useful when measuring a given constant quantity consecutively, in 
conjunction with a range of reference weights defined by upper- and lower-limit values.  

This function can be used in weight mode, counting mode, or percentage mode.  

Limit value input methods 

Either of the following two methods can be used in the different modes: 

(1)  Actual quantity setup method ........ An actual sample is loaded on the balance and its weight 

saved as the limit value. 

(2)  Numeric value setup method ......... The limit value is entered with a key stroke. 

The limit values entered are held in memory, even when the balance is powered down.  

The respective limit values for weight mode, counting mode, and percentage mode are set up 

independently.  

Indication of judgement result 

The [

] mark lights up as either HI, OK, or LO on the left side of the display, indicating the result of 

judgement. 

Judgement Results 

Upper/lower-limit setting 

One-point setting 

HI (excessive) 

Upper-limit value < measurement 
value 

No indication 

OK (appropriate) 

Upper-limit value 

 measurement 

value 

 lower-limit value 

Limit value 

 measurement value 

LO (insufficient) 

Lower-limit value > Measurement 
value  

Limit value > Measurement value 

8.1  Limit Function Setup 

Invoking the function 



  



指を離す

長押し

   



g

Key

released

Continuous

pressing

Press and hold down the  Function  key. 

Release the key when [



] is 

displayed. 
The display changes to the function 
setup and the first item is displayed. 

Содержание AJ-1200E

Страница 1: ...lance AJ E AJH E Series Operation Manual SHINKO DENSHI CO LTD IMPORTANT To ensure safe and proper use of the balance please read this manual carefully After reading this manual store it in a safe plac...

Страница 2: ......

Страница 3: ...ability intrinsic to a tuning fork design help boost operational efficiency Moreover balances with a built in calibration weight AJH E Series can be calibrated by simply turning the calibration knob B...

Страница 4: ...20 8 Limit Function 22 8 1 Limit Function Setup 22 8 2 Setup of Limit Values by Actual Quantity Loads 24 8 3 Setting up Limit Values by Inputting Values 25 9 Calibrating the Balance 27 9 1 Calibratio...

Страница 5: ...ENDED This term indicates steps that the user should take to ensure the quality and reliability of the balance Meanings of Symbols Each symbol is accompanied by an instruction Mandatory Symbol Indicat...

Страница 6: ...ric shock Keep Dry Do not use the balance in a location were it may be subjected to excess moisture Electric shock or short circuiting could occur The balance may be corroded with resultant malfunctio...

Страница 7: ...increase in the balance may lead to inaccurate measurement Unplug Adapter If the balance is to be unused for an extended period of time unplug the adapter This conserves power and prevents deteriorati...

Страница 8: ...le rests in the center of the red circle Adjusters Round pan type Total of two one on either side of the unit s front Square pan type Total of four one on either side at front and rear AC adapter conn...

Страница 9: ...the limit function is active mom Momme M Display of set values from memory If a value is flashing it is being saved CAL Stays on and flashes while span adjustment is in progress Bar graph ct carat oz...

Страница 10: ...t function Function key Brief press toggle switches the units to be displayed in succession g Pcs etc Brief press moves the flashing digit in the setup of a limit value when using the value input meth...

Страница 11: ...the adjusters until the bubble rests in the center of the red circle on the level Round pan type total of two adjusters one on either side at front Square pan type total of four adjusters one on eith...

Страница 12: ...s A few seconds after startup the indication is reset to zero 3 Switching the measurement mode g Pcs Press the Function key Each time the key is pressed the unit of measurement changes as indicated Th...

Страница 13: ...indicate only the weight of the sample loaded in the tare Weighing only the weight of an added sample 4 Resetting the indication to zero 0 T 0 F g Press the Zero Tare key The total weight of the samp...

Страница 14: ...and off it is likely that the balance is being affected by wind or other vibrations Use the windshield or vibration dampers to protect against such adverse effects 0 F g 0 F g Unstable Stable 4 When t...

Страница 15: ...2 Selecting the next item Press the Function key The indication changes to the next item Limit function 3 Selecting an item Pressing the Function key advances the function items at the rate of one ite...

Страница 16: ...udgement Upper limit and lower limit values are set up judges among HI OK and LO Disable Auto zero zero tracking Enable This function automatically sets the zero point exactly to zero to prevent sligh...

Страница 17: ...hether stable Outputs once if stable Outputs if the balance is stable when a sample is loaded after the preceding sample has been removed and the balance indicated zero or less Outputs once if stable...

Страница 18: ...g 5 2 Setup of Units of Measurement When values to are entered prior to use the desired unit of measurement to be displayed can be chosen simply by pressing the Function key For more information on th...

Страница 19: ...are entered in the function items to prior to use the desired unit of measurement to be displayed can be selected simply pressing the Function key For more information on the units of measurement that...

Страница 20: ...Sample unit weight weight unit Sample weight Total g 1 g 500 500 pcs Measured number of samples If samples to be counted deviate widely in weight or a higher measure of accuracy is desired it is reco...

Страница 21: ...ollowing states 1 The weight of one sample measurable unit weight is insufficient For the range of unit weights that can be measured and saved please refer to 14 Specifications on page 41 2 In the sam...

Страница 22: ...ples Pcs M Add approximately an equal number of samples as the number currently displayed Add them gradually in groups of approximately the same number until the number of samples is approximately dou...

Страница 23: ...ce weight 1 Activating the percentage mode Press the Function key to disaplay 2 Setting the reference value Press the Set key The display begins flashing The balance is now ready for reference value s...

Страница 24: ...The weight of the reference sample is insufficient For the limit weight that can be saved limit weight please refer to 14 Specifications on page 41 2 While setting up the reference value in Step 2 th...

Страница 25: ...s the limit value 2 Numeric value setup method The limit value is entered with a key stroke The limit values entered are held in memory even when the balance is powered down The respective limit value...

Страница 26: ...udgement Condition Press the Zero Tare key to select the desired condition 5 Setting the judgement range Judge entire range Judge beyond 5 Press the Function key The display changes to Judgement Range...

Страница 27: ...emory key After the lower limit value has been saved the balance displays it briefly and proceeds to the following setup If One point setup was chosen the setup is complete 4 The upper limit value set...

Страница 28: ...ess the Function key The flashing moves to the digit on the immediate left Each time the key is pressed the flashing digit moves one position left When the leftmost digit is selected the flashing adva...

Страница 29: ...imit value displayed Lower limit value displayed Measurement mode restored Continuous pressing 2 If you make a mistake press the Function key during the setup of actual quantities or the Set key durin...

Страница 30: ...AJ 4200E AJ 6200E AJ 8200E AJ 12KE Span adjustment should be performed with the balance installed perfectly level and without any load on the measurement pan 1 Activating the span adjustment function...

Страница 31: ...y of the balance To implement a calibration as precisely as possible use a weight close to the capacity of the balance Calibration weights can be ordered from Shinko For ordering information please co...

Страница 32: ...elease both fingers at the same time Press the Function key while holding down the Zero Tare key then release both fingers simultaneously Display will change to a flashing and the zero point will be c...

Страница 33: ...er is turned on with the knob in the Calibration position or when set to the Calibration position during measurement 4 Wind or vibration affecting the balance may result in the process not proceeding...

Страница 34: ...2 Calibration of AJH E Series Page 29 GLP Compliant Print Form CALIBRATION MODEL S N ID DATA TIME CAL END NAME ID Title Model of balance Serial No of balance ID No Date of calibration Time of calibra...

Страница 35: ...onnector DIN 5 pin Rear panel Compatible plug TCP 0556 01 0201 Hoshiden supplied with balance Tare subtraction zero adjustment is possible by connecting an external tare subtraction input and a signal...

Страница 36: ...ction with an IBM PC AT compatible Sample connection with PC9801 TXD RXD GND DCD RTS CTS DSR DTR RXD TXD GND 3 4 5 3 2 5 20 7 8 6 4 Balance DIN5P 3 2 5 1 7 8 6 4 TXD RXD GND DCD RTS CTS DSR DTR RXD TX...

Страница 37: ...gic 1 Logic 0 Data bit Start bit Parity bit may be omitted Stop bit 10 4 Output Data By changing the function settings on the main unit of the balance users can select either of the following formats...

Страница 38: ...it may be omitted and replaced with a blank space SP in the lowest order place SP 20H Space zero of leading portion of value leading zero suppress 10 4 4 Units U1 U2 two characters All the codes are A...

Страница 39: ...mands can be transmitted regardless of their data transmission timing 2 When the balance has executed the received command it activates a normal end response or transmits the requested data via the tr...

Страница 40: ...H Outputs once if stable and outputs continuously when unstable Even if the sample is not replaced output of the balance stops when it stabilizes after being output once 7 4FH 37H Pressing Memory key...

Страница 41: ...harge the battery with a weak current when the power is switched on Overcharging will also decrease battery life 2 When the balance is used for the first time after purchase the operating time may be...

Страница 42: ...1 Set a small panel in a large panel An edge with step should face the same side 2 Set the other large panel in the assembly above 3 Set the other small panel in the assembly above 4 Put the cover on...

Страница 43: ...the balance weight range container weight of sample A section of the mechanism is damaged Check the total weight Execute tare subtraction again Contact our Technical Service Division or your local de...

Страница 44: ...lity g 0 01 0 01 0 01 0 01 0 1 0 1 Measurable unit weight in counting mode g 0 01 0 01 0 01 0 01 0 1 0 1 Minimum weight in percentage mode g 1 1 1 1 10 10 Weight measuring method Tuning fork vibration...

Страница 45: ...20 620 820 1200 g 0 001 0 001 0 001 0 001 0 01 0 01 1100 1600 2100 3100 4100 6000 ct 0 01 0 01 0 01 0 01 0 05 0 05 7 7 11 14 21 28 42 oz 0 00005 0 0001 0 0001 0 0001 0 0005 0 0005 0 48 0 70 0 92 1 3 1...

Страница 46: ...05 0 001 0 001 0 001 0 005 0 005 1400 2000 2700 3900 5200 7700 dwt 0 01 0 01 0 01 0 01 0 1 0 1 33000 49000 64000 95000 120000 180000 grain 0 2 0 2 0 2 0 2 2 2 58 85 110 160 210 320 Hong Kong 0 0005 0...

Страница 47: ...5 24 11306 1mom 3 75 18 75 0 13228 0 00827 0 12057 2 41131 1to 11 66380 58 31902 0 41143 0 02571 0 37500 7 5 Unit grain tael Hong Kong tael Singapore Malaysia tael Taiwan momme tola 1g 15 43236 0 0267...

Отзывы: