4

 

98MANAL574   Rev. F 

NSR EXAMPLE 

 

You have Viatran’s Model 574 with a standard range of 0 to 300 PSI.  You would like to set an NSR of 0 to 150 PSI.  
Refer to the Performance Certificate and obtain the following:  

Zero = 0 PSI = 4.00 Ma 

Full Scale = 300 PSI = 20.00 mA 

Cal Value = 90.04 PSI = 8.81 mA  

In order to recalibrate the transmitter, you must determine the cal reading in mA’s at the NSR range.  This is done by  
calculating what the cal output should be at the NSR range.  For the example listed, the standard calibration value is 
set to represent 90.04 PSI.  Once internally set, the calibration (Cal) will always represent this pressure.  
 

Use the formula:  Cal mA’s =  [(16 x Standard Cal Pressure) / desired range] + 4 

 

Substitute Actual Numbers: Cal mA’s = [(16 x 90.04) / 150] + 4 

 

 

 

 

Cal mA’s = (1440.64 / 150) + 4 

 

 

 

 

Cal mA’s = 9.604 + 4 

    Cal 

mA’s 

13.604 

mA’s 

The zero would be set for 4.00 mA’s, and the calibration circuit will read (after adjusting span) 13.604 mA’s when 
activated.  This would provide a new NSR range for the unit of 4-20 mA’s over 0-150 PSI.  

ACCESSORIES - OPTIONAL EQUIPMENT 

 

Polypropylene Conduit Connection Box - NEMA 4X (4 Connection, Dual Terminal Block) .................. 10000PT-25 
Polypropylene Conduit Connection Box (10000PT-25) with 8 Ft. Cable ............................................... 003667.003 
Mounting Bracket (Process Flanges – 2” Pipe Mount) ..........................……………………………......... 003791.003 
Replacement Process Flanges ........................................................................…………………….......... 010454.003 
Digital Indicator (Panel Mount) .....................……………………………….....……............................. Consult Factory 
Direct Mount Indicator ......................………………………………….....................…......................... Consult Factory  

PREVENTATIVE MAINTENANCE

 

Note:

   This checklist should be followed each time a transmitter requires preventative maintenance.  Depending on 

 

product usage cycle, all maintenance should be done on a schedule that is reflective of the application.  

1.  The sensor surface should be visually checked whenever the unit is removed or every 6 months (usage 

dependent).  Any visible damage, especially deep gouges or cuts in the sensor, would require the unit to be 
returned to Viatran for evaluation.  

2.  Care should be used when handling the unit by the connection wires. Continued movement of these wires will 

weaken the seal. Seal failure will result in internal contamination and non-warranty product failure. Visually 
inspect the connection area every 6 months (usage dependent).  

3.  The module door should always remain closed over the adjustment screws. Constant exposure of the adjustment 

screws to mists, dusts or contaminants could cause them to become “frozen” in place, causing the module to 
become locked and non-usable. This would render external adjustment difficult at best. Replacement magnets 
are available if this problem occurs. Hence a visual inspection of the calibration pots should take place on a 2 to 
4 week schedule, or when extreme environmental conditions require it.  

4. 

NEVER 

- Strike a presumed failed transmitter. If the transmitter is in working order, the shock could cause 

additional failure. 

NEVER

 - Remove a transmitter from an instrumentation loop with the power on.  Protection circuitry designed to 

absorb small power surges could become shorted and cause transmitter failure due to “arking”. 

NEVER

 - Ground a transmitter to AC Return (Black).  Instant “fatal” failure occurs. 

NEVER

 - Apply input power to the case ground (Green wire).  Reduced life span of protection components will 

occur.  

MAINTENANCE AND REPAIR

 

All Viatran transmitters have been designed to function free from routine or scheduled maintenance.  Simple cleaning 
of the electrical connector, pressure port threads and pressure cavity on an as needed basis will provide many years 
of satisfactory performance.  Protecting the product from continued exposure to moisture or fluids at the electrical 
connection, breather area (model dependent) will eliminate premature internal failure of the product.  Generally, any 
time the product is removed from service the connector and pressure port threads should be cleaned and the 
pressure cavity flushed with a 316 stainless steel compatible cleaner to prevent media buildup.  

It is suggested that the calibration be verified on a usage dependent schedule.  If the product is in continuous service 
7 days a week, then calibration verification may be necessary every 6 to 8 months.  If the product is in a lab test 
environment a more lenient verification schedule would be appropriate.  In all instances the performance of the 
product will depend on the individual application or process in which it is installed.  More continued usage would 
require a shorter period between calibration verification and product maintenance.  

If a product is perceived to be exhibiting problems, it can be returned to Viatran for analysis and/or repair.  It is 
suggested that only Viatran personnel attempt repair of the product.  Any damage resulting from customer 
disassembly would result in a loss of coverage under the warranty policy.  All Viatran products are able to be repaired 

Содержание 574

Страница 1: ...L574 Rev F INSTALLATION DATA MANUAL MODEL 574 PRESSURE TRANSMITTER 199 Fire Tower Drive Tonawanda NY 14150 Int l 1 716 629 3800 Toll Free 1 800 688 0030 Fax 1 716 693 9162 www viatran com solutions viatran com ...

Страница 2: ...ct our Application Engineers for assistance Remove and retain all instruction manuals and performance certificates that are shipped with the product These documents provide important information on the product s calibration operation safety precautions recommended maintenance re calibration requirements repair service instructions and warranty information These documents are updated from time to t...

Страница 3: ...ble damage 13 To preserve the integrity of the hermetic design the units have no field replaceable or repairable parts BASIC OPERATION INTERNAL CALIBRATION CIRCUIT See Maintenance and Repair for calibration frequency recommendations Note The casting of the module has raised numbers at the adjustment screw positions Those numbers reflect the following function of the assigned screw Screw 1 Top Left...

Страница 4: ... mists dusts or contaminants could cause them to become frozen in place causing the module to become locked and non usable This would render external adjustment difficult at best Replacement magnets are available if this problem occurs Hence a visual inspection of the calibration pots should take place on a 2 to 4 week schedule or when extreme environmental conditions require it 4 NEVER Strike a p...

Страница 5: ...lve 12 months from the date of shipment provided that Viatran s obligation hereunder shall be limited to correcting any defective workmanship and or replacing any defective material f o b factory No allowance will be made for any expense incurred for correcting any defective workmanship and or material without written consent by Viatran This warranty is in lieu of all other warranties expressed or...

Отзывы: