23
Menu
Function Description
modes.
Parameter
Pressure
Set the treatment pressure of the device in CPAP mode, optional
values: 4.0-20.0 cmH
2
O, and the default value is 6.0 cmH
2
O.
Maximum
pressure
Set the maximum treatment pressure of the device in APAP mode,
optional values: 4.0-20.0 cmH
2
O, and the default value is 12.0
cmH
2
O.
Note: The pressure value must be greater than the current
minimum pressure value.
Minimum
pressure
Set the minimum treatment pressure of the device in APAP mode,
optional values: 4.0-20.0 cmH
2
O, and the default value is 4.0
cmH
2
O.
Suction
pressure
Set the suction pressure of the device in S, S/T, and T modes. The
optional values are 6.0-25.0 cmH
2
O, and the default value is 10.0
cmH
2
O.
Note: The pressure value must be greater than the current
expiratory pressure value.
Expiratory
pressure
Set the expiratory pressure of the device in S, S/T, and T modes.
The available values are 4.0-25.0 cmH
2
O, and the default value is
6.0 cmH
2
O.
Inspiration
time
Set the inspiratory time of the device in S/T and T mode, optional
value: 0.3-4.0 seconds, the default value is 1.0 second.
Respiratory
rate
Set the inspiratory time of the device in S/T and T modes,
optional values: 5-30 bpm, and the default value is 12 bpm.
Pressure rise
time
Set the pressure rise time of the device in S, S/T, and T modes,
optional values: 100--900 ms, and the default value is 200 ms.
Inspiratory
trigger
sensitivity
Set the inspiratory trigger sensitivity of the device in S, S/T mode,
optional values: automatic, 1-5 gears, and the default value is 3
gears.
Exhalation
trigger
Set the exhalation trigger sensitivity of the device in S, S/T mode,
optional values: automatic, 1-5 gears, and the default value is 3