20
Menu
Function Description
Humidify
Humidification
grade
Set the humidification level of the humidifier, optional values:
automatic, 0-5 gears;
Level 0 is to turn off the humidification function. The greater the
number of levels, the greater the humidification capacity. The
default is 0 level.
Comfort
Expiratory
hypotension
Set the level of expiratory pressure reduction, and automatically
reduce the pressure in the mask according to the breathing rhythm
during the expiration phase to improve comfort. Optional values:
0-3 gears, 0 gear is to turn off this function, default value: 2 gears.
Automatic start Turn on/off the automatic ventilation function; when turned on,
the device will automatically start ventilation after the user wears
the mask and breathes; Optional value: on/off, the default value is
on.
Automatic stop Turn on/off the automatic stop ventilation function; when turned
on, the device will automatically stop the ventilation after
detecting that the user takes off the mask; Optional value: on/off,
the default value is on.
Delay start
pressure
Set the starting pressure of the delay boost function. Optional
value 3-20cmH2O,
Default value: 4 cmH2O.
Note: The initial pressure value must be less than the set pressure
value.
Delay boost
time
Set the delay time of the delay boost function, optional value:
0-60 minutes, when set to 0, the delay boost function will be
closed, default value: 15 minutes.
Preheat
Turn on/off the pre-heating function. After turning on, the
humidifier will start to work for 30 minutes before being
ventilated to pre-heat.
Report
View usage
information
Display the usage of the device, including usage time, sleep
quality information, etc.