![Viasensor G200 Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/viasensor/g200/g200_operating-manual_1011568052.webp)
OMG200N1.24
G200 Analyzer Range
VIASENSOR
Page
52
G210
TECHNICAL SPECIFICATION
GAS RANGES (continued)
Measurement
accuracy
CO
Resolution:
1ppm
Accuracy:
±5.0% of range @ constant
temperature and pressure
Temperature dependence:
Zero reading
±5ppm over operating temperature range
Pressure dependence:
TBA
Response time, T
90
CO
2
£
20 seconds
O
2
£
60 seconds
1
Conditions during factory calibration, typically 68
of
F, 401.5” W.C.
FACILITIES
Visual and audible alarm
User selectable N
2
O, CO, CO
2
and O
2
alarm levels
Communications
USB type B mini-connector, HID device class
Data storage
1000 reading sets + 270 events
PUMP
Flow
100cc/min typically (300cc/min MAX)
ENVIRONMENTAL CONDITIONS
Operating temperature
41°F to 104
°
F
Operating pressure
200.7” W.C. to 602.2” W.C.
Relative humidity
5% to 95% non condensing
IP rating
IP40
PHYSICAL
Weight
1.1 lbs.
Size
6.5”L, 3.93”W, 2.17”D
Case material
ABS / Polypropylene with silicone rubber inserts
Keys
17 resin capped silicone rubber keys
Display
Liquid crystal display, 128 x 64 pixel
With RGB LED back-light
Gas sample filters
User replaceable PTFE water trap filter
?
Note:
Due to VIASENSOR’s continuous program of improvement, this specification is
subject to change without prior notice.
Содержание G200
Страница 4: ......
Страница 42: ...OMG200N1 24 G200 Analyzer Range VIASENSOR Page 42 ...