OMG200N1.24
G200 Analyzer Range
VIASENSOR
Page
10
pressure of the gas needs to be reduced before it is passed to the instrument. This
reduces the pressure to 40.15” W.C. suitable for the instrument. It is recommended that
the operator uses the pressure regulator kit supplied by VIASENSOR.
Warning
Do NOT use the instrument at full piped pressure (maximum
pressure 120.44” W.C.).
3.1.5 Soda Lime CO
2
Filter
The G200 instrument must only be used with a soda lime CO
2
filter attached. The filter
removes CO
2
if present when a reading is taken.
This filter should always be fitted when using a G200 instrument as any CO
2
will cross-
contaminate the N
2
O reading. For further information please refer to section ‘8.5 – Cross
Gas Effects’.
The life of the filter will vary depending on usage. The crystals change color from pale
green-blue to white-grey. The filter element should be discarded when a significant
percentage of the crystals start to turn a white-grey color see below:-
Assemble Filter and Connect to Instrument as Shown
New Filter (Pale Green-Blue)
Used Filter (White-Grey)
The filters are supplied in a sealed storage bag to
improve their shelf life. The parts supplied should be
assembled as per the diagram shown. Please note the
sample filter is still in place.
?
Note: A soda lime CO
2
filter can also be used to
zero other G Range instruments.