GLOSSMETER

1.  Feasures:

Designed and manufactured in accordance 
with international standard ASTMD523, 
ASTMD1455, ASTMC346, ASTMC584, 
ASTMD2457, DIN ENISO2813, DIN67530, 
ENISO7668, JISZ8741, MFT30064, 
TAPPIT480, GB9754, GB/T13891, GB7706 
and GB8807. Its technical parameters 
conform with JJG696-2002. Flexible use 
with its separate probe.

2.  Applications: 

Widely used in Floor maintenance, Surface 
cleaning quality control, Stone and tile gloss 
measurement, Checking printed matter, 
Quality control of paint and ink, Polished 
metal surface , measurement (chrome 
plating), Inspection of paint protection and 
waxing, Auto-body paint inspection, Surface 
inspection of plastic moldings, Evaluation of 
detergents and washers,Checking masonry
and building exteriors.

3.  Technical Specifications

Measuring geometry: 
        GM-6: 60° 
        GM-26: 20° & 60° 
        GM-268: 20°, 60° & 85°

Range: 0.1 to 200 gloss units
Accuracy: ±1.0 gloss unit (against reference 
standard )
Resolution:  0.1 gloss unit 
Repeatability: ± 0.5 GU  (0 ... 99.9)
Measuring area: 7x14mm ellipse
Data memorized:  254 groups
Data output: RS 232C serial interface 
Working condition:
       Temperature range: 0°C- 40°C
       Humidity: up to 85%
Power supply:  Lithium Battery
Size: 140x45x75 mm
Weight: 305 g

4.  Accessories

        Carrying case.............................1pc
        Operation manual.......................1pc.
        Power adapter............................1pc.
        Holder with calibration standard...1pc.
        Optic cleaning cloth….................1pc.

5.  Optional accessories

       Cable & software for RS232C.
       USB adaptor. 
       Bluetooth interface.

3-1

3-2 3-3 3-4

3-5 3-6 3-7

6.  Front Panel Descriptions

3-1 Display
3-2 Measuring key
3-3 Delete key 
3-4 Read key
3-5 Power key
3-6 Angle conversion key
3-7 Calibration key

7.  Power ON / OFF

Press the key to power on the meter. To 
turn it off, just press this key again. This 
meter is powered with Lithium Battery. 
Use the battery adapter for charging. 

8.  Calibration

1.Press the “ANGLE key” to select the 
vight measuring geometry
2.Lay the main unit into the holder with 
calibration standard of high gloss black 
tile. Take a reading by pressing the 
“MEAS key.” Compare the measured value 
with the value marked on the tile. If it 
matches the value on the tile holder, the 
instrument is within calibration and ready 
for use. If the measured value does not 
match the assigned value, just press the 
“CAL key.”
It is necessary to inspect the condition of 
the calibration tile and instrument optics 
before each calibration. Any dust or debris 
on the optic should be blown from the 
lenses using 

dry clean air, the optics must not be 
touched. If there are any permanent marks 
or scratches on the lenses, the instrument 
is no longer suitable for measuring and 
should be returned to an authorized 
service center. The calibration tile must be 
perfectly clean from smears and scratches 
before attempting calibration. Fingerprints 
and dust can be removed with the supplied 
optic cleaning cloth.

9.  Measurement

Remove the holder and lay the 
Measurement Aperture just against the 
position that you want to measure. Take a 
measurement by pressing the “MEAS 
key.” The reading on the Display Screen 
within 2 seconds is the gloss value of that 
position. Meanwhile the result is stored 
into memory automatically.
When the instrument is placed on a 
sample the aperture is hidden, the centre 
of the measurement area can be 
pinpointed by the intersection of the 
arrows marked on the front of the 
instrument case with those on the side.

10.  Data Memorizing

When taking measurements, all values will 
be stored automatically in memory and the 
number of stored reading is accordingly 
increased by 
10.1 The gloss meter can store up to 254  g
roups of data with measurement conditions. 
If the memory is full, the tester will auto save 
the new reading and discard the oldest one. 
That means the tester only hold the last 254 
groups of data in memory.

11.

  

Memory Recall

Under the measuring mode marked by ‘M’ 
on the display , press the “RD key” to enter 
into mode of viewing stored values marked 
by “R”. Press “RD key” or “CAL key” to recall 
stored values forwards and backwards. To 
return to the measurement state, just press 
any key otherthan “RD key” or “CAL key”.

12.  Delete data in me mory 

In the measurement state marked by “M” on 
the display , the new reading can be deleted 
by pressing “DEL key”, and the number of 
stored reading is accordingly decreased by 
1. To clear memory, just press and hold 
“DEL key” for 3 seconds till the number of 