HOW TO CHARGE THE INTERNAL BATTERY

  

1.  Connect the charger (AC Adapter 12 VDC, 800mA) to the scale, and then plug the charger 

into an AC outlet. Make sure that the AC voltage appearing at the wall outlet matches 
the input voltage marked on the AC adapter.
 

2.  After  the  charging  period  expires,  unplug  the  charger  from  the  AC  outlet,  then  from  the 

scale. The scale is now ready for use under its own battery power.  

NOTE: The charger may be left connected to the scale indefinitely without damage to the in-
ternal battery.  

HOW LONG TO CHARGE THE INTERNAL BATTERY

  

In general, the battery should be allowed to charge a minimum of 1.5 hours for every hour of 
use. If you discharge the battery below 50% and do not allow the proper time for charging, you 
may start to notice a decline in the usage period. This is normal and eventually the battery must 
be replaced.   

REPLACING THE BATTERY

  

The recommended practice when removing the battery is to disconnect the ground connection 
(black) first, then the red terminal. This ensures that a short circuit will not occur from a battery 
lead or fuse lead touching the grounded housing while disconnecting the other terminal. Similar-
ly, the ground should be connected last when installing a new battery. 

The US government has classified the internal battery as hazardous waste. Do not

 

place battery 

into landfill. An automotive store or a local waste agency may accept the batteries for recycling. 
Contact the manufacturer for more information.  

Operation – weigh beams

  

Each beam contains two buttons and one LED.  

 

The BLUE button is used to power up the beams 

 

The RED button is used to power down the beams  

The beams have several operating modes to save battery life and also to alert the user when its 
time to replace the batteries. You can determine the operating mode by observing the blinking 
behavior of the blue LED contained within each beam:  

LED Functionality – Blinking interval 
During Power-up (Initialization) 

ON for 5 seconds 

   

FULL BATTERY 

Blink interval  

Working Mode 

1.5 seconds  

Sleep Mode (power savings) 

4 seconds  

   

LOW BATTERY 

Blink interval  

Working Mode 

10 seconds 

Double Flash 

Sleep Mode 

10 seconds 

Double Flash 

 

Please replace the batteries when the blink interval is 10 seconds!  

NOTE 1: Beams will shut down completely after two hours of inactivity. 
NOTE: 2: Beams will only enter sleep mode when the indicator (display) is OFF. 

Содержание WB5040-SWP

Страница 1: ...MODEL WIRELESS PORTABLE SCALE User Manual Manual_WB_BT_V1 ...

Страница 2: ...t NOTES The indicator comes with a pre installed rechargeable battery The weigh beams come with pre installed C cell alkaline batteries Do not mix the type of batteries Replace all batteries at the same time If the scale is not to be used for a long period of time remove all batteries from the battery compart ment to avoid leakage which may cause damage to the instrument To extend the life of your...

Страница 3: ...nternal lead acid rechargeable battery Before using the indicator for the first time please charge the battery overnight The indicator s battery should operate for about 85 hours if left on continuously The display unit is configured to power down after 30 minutes of inactivity The battery can be charged while ON or OFF and the indicator can be operated while it s charg ing unless the state of cha...

Страница 4: ...ead touching the grounded housing while disconnecting the other terminal Similar ly the ground should be connected last when installing a new battery The US government has classified the internal battery as hazardous waste Do not place battery into landfill An automotive store or a local waste agency may accept the batteries for recycling Contact the manufacturer for more information Operation wei...

Страница 5: ...l alkaline batteries 1 Turn the bean OFF Press the RED button 2 Remove the rubber cover 3 Grasp the metal cylinder inside and carefully pull free from the housing 4 Unscrew either end cap to reveal batteries 5 Exchange the batteries be careful to note polarity 6 Replace the end cap 7 Slide the metal cylinder back into the metal housing being careful not to crimp or cut the wire Keyboard and Displa...

Страница 6: ... below The display unit can show overall system battery health See the Error Messages for more in formation Function Keys Units This key toggles the indicator between lb and kg Zero Off This key sets the indicator to display zero weight provided the following conditions are met 1 The indicator is displaying Gross weight 2 The scale is not in motion 3 The scale is not in overload see error codes Pr...

Страница 7: ...trying to troubleshoot setup problems 4 Remove all items from the beams and press the ZERO OFF key to zero out the displayed value 5 Press the NET GROSS key to save the zero point value The display will show EndC0 momentarily and then revert back up to F16 6 Press the ON PRINT key to progress to the F17 menu 7 While at the F 17 screen scroll down once using the ZERO OFF key to enter span calibra t...

Страница 8: ...stablish a proper calibration Most commons causes include incorrect load cell wiring a mechanical obstruction or a faulty damaged load cell Error Messages Indicates that the weighing capacity of the scale has been exceeded The weight limit of the WB5040 SWP is 5000 pounds Indicates that the weight on the scale is negative Press the Zero Off key Beam 1 not found ensure that beam 1 is po wered ON an...

Страница 9: ...ion MAIN COMPONENTS TS 500SS RF stainless steel digital indicator and AC adapter Two 40 weigh beams BEAM CONSTRUCTION Carbon steel tube High grade epoxy paint finish LOAD CELLS 350 ohm two per weigh beam Overload protection 200 Operating temperature 14 F to 104 F 10 C to 40 C NTEP approved Digital Indicator Specifications NEMA 4X washdown stainless steel enclosure Operating temperature 14 F to 104...

Страница 10: ...m that is displaying the lower weight Check underneath the weigh beam for any obstructions Make sure that the weigh beam feet are not screwed in so far as to restrict downward movement of the weigh beam Indicator displays six small zeros Scale is overloaded Remove weight from scale Scale turns off on its own The indicator has a power conservation feature set to automatically power off the scale af...

Отзывы: