Specifications and displayed error codes. 

Model 

Capacity  / Graduation 

Dimension mm 

Battery 

Own weight 

VB-10K 

Max 10000 daN / 2 daN 

260 x 210 x 54 

1 pcs alkaline 9 V 

3,2 kg 

OPERATOR INTERFACE

 

Display  

0.75" (19 mm) 7-segment, Liquid Crystal, 6-Digit 

Additional Symbols 

Net, Gross, Stable, Tare, lb, kg, Zero 

Keyboard 

5-key flat membrane panel

ENVIRONMENTAL

 

Operating Temperature 

-10 to +40 grad C

Storage Temperature 

-25 to +70 grad C

Displayed error codes.

 

CODE 

MODE 

MEANING / POSSIBLE SOLUTION 

Normal Operating 
Mode 

Gross  Overload.  A  force  greater  than  the  rated  capacity  has  been  applied  to the Force Meter. Remove the 
force from the platter or try re-calibrating the Force Meter. Otherwise, check for a bad load cell connection 
or possible load cell damage due to overloading. 

bAtt 

Normal Operating 
Mode 

Indicates a low battery condition. 

Err 0 

Span Calibration 
Mode (F17) 

Keyed-in  force  value  is  larger  than  full-Force  Meter  capacity.  Use  a  smaller  test  force  or  check  keyed-in 
value. 

Err 1 

Span Calibration 
Mode (F17) 

Keyed-in force value is less than 1% of full-Force Meter capacity. Use a larger test force or check keyed-in 
value. 

Err 2 

Span Calibration 
Mode (F17) 

There is not enough load cell signal to produce the internal counts necessary to properly calibrate the Force 
Meter.  First  check  all  load  connections.  Use  F16 mode to view internal counts. See Appendix C for more 
information.  

Err 3 

All Modes 

Non-volatile memory read error. One or more setup parameters have been lost. 

Err 4 

All Modes 

Non-volatile memory writes error. The unit needs service. 

Err 5 

Key-in Span 
Calibration Mode 
(F20) 

You have attempted to enter a zero value for C1. Enter a known calibration value greater than zero. 

Err 7 

Initialization 

No reading from the ADC. Make sure there is a load cell(s) connected to the indicator at start-up. 

Err 9 

Normal Operating 
Mode 

Span calibration value has been lost. Re-calibrate the Force Meter. 

Hantverksvägen 15, 76493 Väddö Sweden

Tel  +46 176 208 920

 

 

info@vetek.

com

 

 www.vetek.

com

Содержание VB2-10K

Страница 1: ...Normally the only instruction shipped together with the Force Meter Alkalinebattery9V Calibrationswitch The Weighingblock is a robust Force Meter It is designed for portable use When transported by car between different working places we recommend using of a protected space or a case This will extend the interval for service ON PRINT NET GROSS OFF ZERO TARE O 9VDCjack 2 1mmcenter minus Manual_VB2 ...

Страница 2: ...gh the User menu The configuration section of the manual explains how to use the five front panel keys to maneuver and save settings in both menus The Force Meter is equipped with 9 V alkaline battery If the battery is not good enough to ensure a correct value the Force Meter will turn off Keyboard functions The calibration switch besides the battery is only for calibration Indicator ON Normallydi...

Страница 3: ...6 To move to the selection value press ZERO once 7 Increase the flashing digit bypressing lb kg Decrease the flashing digit bypressing ZERO 8 To save the value press NET 9 Press O to go back to parameter number 10 To go back to normal menu press Calibration Switch one time for normal mode NOTES ON THE SETUP MENU There is an F21 sub menu present that is for FACTORY USE ONLY DISPLAY LCD Enunciator M...

Страница 4: ... a very fast response is needed 1 2 4 8 F7 Overload Limit Selects the desired formula which determines the point at which the indicator shows overload All selections are based on the primary unit selected in F8 FS Full Force Meter in primary units FS FS 2 FS 1d FS 9d F8 Calib Unit Selects the primary base unit to be used in the calibration process Also the default unit for normal operation 1 prima...

Страница 5: ...pan calibration procedure 1 Scroll to F 17 then press ZERO to enter span calibration menu 2 The display will momentarily show C 1 for the span calibration followed by a value with one flashing digit This value will be zero Place the test force on the Force Meter 3 Pressing TARE or ON will change the position of the flashing digit 4 Increase the flashing digit by pressing O Decrease the flashing di...

Страница 6: ...rger than full Force Meter capacity Use a smaller test force or check keyed in value Err 1 Span Calibration Mode F17 Keyed in force value is less than 1 of full Force Meter capacity Use a larger test force or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to produce the internal counts necessary to properly calibrate the Force Meter First check all load c...

Страница 7: ... 2 Reset to normal mode 3 Press OFF and ON Check the Loadcell 1 Check the Loadcell with a universal volt instrument see encl table The pin close to the corner is number 1 Pin Description 1 Out 5VDC 2 In 3 Out 0 VDC 4 In If some parameters by mistake has been changed 1 If only the first F1 has been changed Set back to right value in accordance to the table 2 If many parameters has been changed Chec...

Страница 8: ...fnot wait a fewseconds until it appears Handling 1 Press ON to start the Force Meter 2 If necessary press ZERO to obtain a weight force reading of zero 3 Press OFF to switch off the Force Meter Hold 3 seconds 4 The Force Meter switches off automatically after 20 minutes when not in use Force measuring 1 Unload Force Meter and press ZERO 2 Place the object on the Force Meter s platter and allow the...

Отзывы: