page 6
Specifications and displayed error codes.
Model
Capacity / Graduation
Dimension mm
Battery
Own weight
VB-10K
Max 10000 daN / 2 daN
260 x 210 x 54
1 pcs alkaline 9 V
3,2 kg
OPERATOR INTERFACE
Display
0.75" (19 mm) 7-segment, Liquid Crystal, 6-Digit
Additional Symbols
Net, Gross, Stable, Tare, lb, kg, Zero
Keyboard
5-key flat membrane panel
ENVIRONMENTAL
Operating Temperature
-10 to +40 grad C
Storage Temperature
-25 to +70 grad C
Displayed error codes.
CODE
MODE
MEANING / POSSIBLE SOLUTION
Normal Operating
Mode
Gross Overload. A force greater than the rated capacity has been applied to the Force Meter. Remove the
force from the platter or try re-calibrating the Force Meter. Otherwise, check for a bad load cell connection
or possible load cell damage due to overloading.
bAtt
Normal Operating
Mode
Indicates a low battery condition.
Err 0
Span Calibration
Mode (F17)
Keyed-in force value is larger than full-Force Meter capacity. Use a smaller test force or check keyed-in
value.
Err 1
Span Calibration
Mode (F17)
Keyed-in force value is less than 1% of full-Force Meter capacity. Use a larger test force or check keyed-in
value.
Err 2
Span Calibration
Mode (F17)
There is not enough load cell signal to produce the internal counts necessary to properly calibrate the Force
Meter. First check all load connections. Use F16 mode to view internal counts. See Appendix C for more
information.
Err 3
All Modes
Non-volatile memory read error. One or more setup parameters have been lost.
Err 4
All Modes
Non-volatile memory writes error. The unit needs service.
Err 5
Key-in Span
Calibration Mode
(F20)
You have attempted to enter a zero value for C1. Enter a known calibration value greater than zero.
Err 7
Initialization
No reading from the ADC. Make sure there is a load cell(s) connected to the indicator at start-up.
Err 9
Normal Operating
Mode
Span calibration value has been lost. Re-calibrate the Force Meter.