page 4
PARAMETER
DESCRIPTION
CODE/VALUE
F1
Graduations
Specifies number of the Weighingblocks graduations. Value should be consistent
with legal requirements and environmental limits on the useful system resolution.
500
1,000
1,500
2,000
2,500 3,000
4,000
5,000
6,000 8,000
10,000
F2
Span Gain
Span Gain is related to A/D integration time. The larger the span gain, the higher the
internal resolution, but the slower the update speed.
25
50
75
100
150
200
F3
Zero Track Band
Selects the range within which the Force Meter will automatically zero. Note that the
Force Meter must be in standstill to automatically zero. Selections are in Display
Divisions. “d” = graduation
0d
0.5d
1d
3d
5d
F4
Zero Range
Selects the range within which the Force Meter may be zeroed. Note that the indicator
must be in standstill to zero the Force Meter.
100%
1.9%
2%
20%
F5
Motion Band
Sets the level at which motion is detected by comparing the present display update
with the previous one. If motion is not detected for two seconds or more, Force Meter
is in standstill and can process a Print or Zero command. Maximum value varies
depending on local regulations.
0,25
1d
3d
5d
10d
F6
Digital Filter
Averages weight readings to produce higher accuracy. The higher the filter number,
the greater the accuracy but the slower the response time. Choose 4 or 8 unless a very
fast response is needed.
1
2
4
8
F7
Overload Limit
Selects the desired formula which determines the point at which the indicator shows
overload ("
"). All selections are based on the primary unit selected in F8.
"FS" = Full Force Meter in primary units.
FS
FS + 2%
FS + 1d
FS + 9d
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process. Also the default
unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1
2
F9
Display Div.
Determines the desired weight increments. Value should be consistent with legal
requirements.
1
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0
0,0
0,00
0,000
0,0000
F16
Zero Calibration
Places indicator into the zero calibration routine. Scrolling down with the
ZERO
key
one level begins the procedure.
Press
ZERO
key to begin sequence
F17
Span Calibration
Places indicator into the span calibration routine. Scrolling down with the
ZERO
key
one level begins the procedure.
Press
ZERO
key to begin sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to the default
settings. USE WITH CAUTION!
Press the
ZERO
key twice to execute.
A5
Disable the lb/kg
Key
Allows the lb/kg key to be disabled so that an operator cannot accidentally press the
key and change the displayed units.
"0" = Disable the lb/kg key "1" = Enable the lb/kg key
0
1
A10
Auto Power Off
Period
Selects the auto off time period in minutes:
“Off” = Disabled (Always ON)
1, 2,
4
, 10, 20
,
60, off