Press[MODE] 

dFULt

 

Company default setting 

[ENTER] 

S dFL

 

Save default setting value 

   [MODE] 

dFULt

 

After press [mode] key, DYNALINK save data and 
return back to normal status

handle instrument return 

back to menu status 

 Press[MODE] 

to select 

  ESC

 

  [ENTER] 

  XXX

 

handle instrument return back to normal 
communication status 

2

 LOCAL Setting 

After Remote setting ,if DYNALINK RF parameters are changed, the handle instrument will can not communicate with 
DYNALINK, it must be set the parameters same as DYNALINK. setting method as following: 

KEY 

DISPLAY 

ILLUSTRATION 

[ZERO][TARE] 

[ON/OFF] 

00000

 

After display

00000

can release key 

88888

 

Display Twice 

 Ert 

 

Display Twice 

u 3.2

 

Display Current handle instrument Software version 

U X.XX

 

Display Current handle instrument voltage 

LOCAL

 

Local parameter setting 

[MODE] 

CH E3

 

Display current wireless channel(default is E3)Can be 
selected from 00 to FF, press [ZERO] to adjust digit, 
press [TARE]to move digit 

[MODE] 

   

AddrH

 

Address code high 4bits,Used for wireless net, sole 
instrument don’t adjust 

[MODE] 

00CCC

 

press [ZERO] to adjust digit, 
press [TARE]to move digit 

[MODE] 

   

AddrL

 

Address code low 4bits,Used for wireless net, sole 
instrument don’t adjust 

[MODE] 

CCCCC

 

press [ZERO] to adjust digit, 
press [TARE]to move digit 

[MODE] 

pAirH

 

Pair code high 4bits,Used for wireless net, sole 
instrument don’t adjust  

[MODE] 

H0000

 

press [ZERO] to adjust digit, 
press [TARE]to move digit 

[MODE] 

pAirL

 

Pair code low 4bits,Used for wireless net, sole 
instrument don’t adjust 

Содержание ERT DYNALINK

Страница 1: ...al Nordens största webbshop för vågar Hantverksvägen 15 76493 Väddö Sverige Tel 46 0 176 208920 Web www vetek se Email info vetek se TECHNICAL MANUAL ERT DYNALINK With optional remote display Technical_Manual_ERT_V2 ...

Страница 2: ...econdS 00000 After display 00000 can release key 88888 Display Twice Ert Display Twice u 3 2 Display Current DYNALINK Software version U X XX Display Current DYNALINK DC voltage SEtdC DC voltage calibration Press ENTER key start DC voltage calibration procedure Press ENTER U AdJ DYNALINK and handle instrument Display synchronization ENTER U X XX Display voltage value press ZERO to adjust digit pre...

Страница 3: ...earity Press MODE SEt3P start to 3 points calibration procedure ENTER CAL 3 ENTER CALSP Zero position confirmation ENTER LoAd1 First time load ENTER XXX Input first time load value press ZERO to adjust digit press TARE to move digit ENTER LoAd2 Second time load value need more than first time value 20 ENTER XXX Input second time load value press ZERO to adjust digit press TARE to move digit ENTER ...

Страница 4: ...ease key 88888 Display Twice Ert Display Twice u 3 2 Display Current handle instrument Software version U X XX Display Current handle instrument voltage LOCAL Local parameter setting MODE CH E3 Display current wireless channel default is E3 Can be selected from 00 to FF press ZERO to adjust digit press TARE to move digit MODE AddrH Address code high 4bits Used for wireless net sole instrument don ...

Страница 5: ...ompany s default setting Just press PEAK and ENTER key 3 seconds 3 Recall handle instrument default value If the handle instrument s parameter were changed by wrong operation User can recall company s default value Switch off handle instrument press TARE ENTER and ON OFF key 3seconds handle instrument will display L dFL and automatic power OFF Press ON OFF key default value will be load into handl...

Отзывы: