DGT / DGT60 / DGTQ / DGTP
28
from 3) >> 16).
20)
Saving the Calibration
Once the calibration has been made of all the necessary points, press many times the
C - ON/OFF
key until
the indicator shows "SAVE? in the display: confirm with
to store and return to weighing.
N.B.
: the calibration points must be in increasing order (point 1 < point 2 < point 3).
4.2 Calibration procedure for DEPENDENT CHANNELS MODE (could eventually be digitally
equalized)
Premise:
this procedure is to be followed in case one has to calibrate a scale having 2, 3 or 4 load cells in the
"DEP.CH" functioning mode, (
see section. 3.2: tYPE –Application type
) and wants to equalize directly from the
indicator, without using the external junction boxes.
In this case one should connect each single cell to one of the available channels of the instrument.
Note:
it is always better to carry out the
equalization procedure
(steps from 4 to 9), but it isn’t compulsory (in some
applications, it can not be done).
Calib (*)
ConFiG (§)
GrAV. (*)
9,75001 … 9,84999
diV
dECi
u.M.
1, 2, 5, 10, 20, 50
G, Lb, t, kG
CALib.P
1, 2, 3, no
rAnGE 1
rAnGE 2
tP 0
ddt 1
tp 1
ddt 2 (§)
n tP
EquAL (§)
tp 2 (§)
ddt 3 (§)
tp 3 (§)
0.CALib(*)
SEtuP
nChan (*)
1..4
...
SETUP
ENVIRONMENT
tYPE (*) (§)
dEP.Ch
CEL.SEn
CEL.CAP
dEAd.Ld
Kno.WGt
thEo.CA
(§) =
indicates that the parameter is visible only in certain programming conditions; see the explanation of the parameter itself for
the details.
(*) =
indicates that with an approved instrument the parameter can be not visible or read only; see the explanation of the
parameter itself for the details.
1)
Enter in the SET-UP ENVIRONMENT of the indicator (when turned on, press for an instant the
TARE
key
while the firmware version).
2)
Select the type of application:
Enter in the "tyPE" step and select the "dEP.Ch" parameter.
3)
Select the number of connected cells
Select the number of cells (in other words, the number of the channels, from 2 to 4) in the
SEtuP >>
ConFiG >> nChan
step and press
ENTER/PRINT
.
4)
Repeat the passages from 4) to 10) described previously in section 4.1 "Calibration procedure for
INDEPENDENT CHANNELS’ mode"
Содержание DGT
Страница 1: ...WEIGHT INDICATORS DGT DGT60 DGTQ and DGTP Technical Manual DGT_DGT60_DGTQ_DGTP_06 01_09 11_EN_T...
Страница 65: ...DGT DGT60 DGTQ DGTP 65 11 CONNECTION SCHEMES 11 1 DGT4 DGT60 DGT4 AN MOTHERBOARD I O BOOT J 1...
Страница 68: ...DGT DGT60 DGTQ DGTP 68 11 2 DGT4 PB MOTHERBOARD I O BOOT J 1...
Страница 71: ...DGT DGT60 DGTQ DGTP 71 11 3 DGTQ DGTQ AN BOARDS...
Страница 72: ...DGT DGT60 DGTQ DGTP 72 I O BOOT J1 J2 CAL...
Страница 76: ...DGT DGT60 DGTQ DGTP 76 11 4 DGTQ PB BOARDS...
Страница 77: ...DGT DGT60 DGTQ DGTP 77...
Страница 81: ...DGT DGT60 DGTQ DGTP 81 11 5 DGTP DGTP AN DGTP PB MOTHERBOARD...