2021/12/18 07:41
19/19
Laser Controller
Product Manuals - https://www.vescent.com/manuals/
If the Temperature with the above TIME settings begins to run away, then quickly turn the
“PROPGAIN” Potentiometer CCW until it stabilizes.
Spectroscopy-Certified
TM
DBR lasers.
Sub-mK stability requires a proper thermal design and proper tuning of the temperature controller to
the thermal plant. If you did not purchase a D2-100 Diode Laser with your Laser Controller, please
read the section on tuning the temperature controller.
Libbrecht and Hall, A Low-Noise, High-Speed Current Controller, Rev. Sci. Inst. 64, pp. 2133-2135
(1993).
Best noise performance obtained with Laser Controller powered by D2-005 placed 5 ft away from
Laser Controller. Placing D2-005 closer to Laser Controller may increase 60 (50) Hz noise and
harmonics. Specifications not guaranteed when Laser Controller is powered by device other than
D2-005, or when placed near other noise sources.
Laser module temperature changed while Laser Controller module held fixed.
Laser module temperature fixed while Laser Controller temperature change.
Temperature at which thermistor reads 10 kΩ
,
1mA resolution with D2-105-500
Courtesy of
From:
https://www.vescent.com/manuals/
- Product Manuals
Permanent link:
https://www.vescent.com/manuals/doku.php?id=d2:laser_controller
Last update: 2021/12/17 00:00