JETTING THE NANOSONDE WITH COMPRESSED AIR
When using the device with compressed air always refer to the user manual of the
jetting unit and follow safe work practices.
Use of compressor can lead to the NanoSonde exiting the duct at a great
speed.
Precaution: Always use a catcher device at the end of the duct. If changing the
direction of the jetting due to the device getting stuck, remember to move the
catcher device to the other end of the duct. Do not jet the NanoSonde with
pressure exceeding 4 bars. Always use safety goggles, when using the NanoSonde
with compressed air.
MAINTENANCE
CLEANING
Clean with clean soft cloth. Do not use abrasive agents or solvents for cleaning.
Take care to avoid dirt getting into the unit. Regularly inspect O-ring and replace if
damaged. If water gets into the battery compartment, allow it to dry at room
temperature.
STORAGE
Store inside storage box. Respect the temperature limits stated in the technical
data
when storing the device. Always remove the batteries from the product
during storage to avoid leakage. After a long time of storage, check the accuracy
of the device.
TROUBLESHOOTING
Problem
Possible cause
Remedy
NanoSonde does not
turn on.
Battery is weak.
Sonde is damaged
Replace battery. There
are no serviceable parts
in the device. Contact
your distributor or
manufacturer
NanoSonde
unintentionally turns off
Battery is weak
Replace battery
No signal is detected
Battery is weak
Device is used in
conductive duct
Replace battery
Use another Sonde