![Vertysystem Apolix Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/vertysystem/apolix/apolix_instructions-for-use-manual_1008464006.webp)
Apolix
Electronic and automatic polymerizer
Rev. 10/2019
Instruction for use
a.gree srl
Via Retrone 14, Altavilla Vicentina 36077 (VI) Italy
+39 0444 960 991 [email protected]
6/7
vertysystem.com
LEGEND
TEMPERATURE (45°C)
SINGLE BEEP
POLYMERIZATION
DOUBLE BEEP
END OF THE CYCLE
AUTOMATIC AIR
DISCHARGE COMPLETED
DURATION
30 min.
10 min.
TOT.
40 min.
P.0 | VERTYS PROTHESIS EXCEL
Ideal programme for cold curing resin
for full denture resin, charged with
composite micro-pearls. For a very
high quality polymerization set the
pressure to 2 Bar and fill the chamber
with at least 3/4 of water.
III. ONLY PRESSURE SETTING
1. Turn on the machine using the ON/OFF switch (2)
2. Hold the two arrows (e) at the same time for 3 seconds. The display (f) will say “AIR”
3. The pressure will be the same as the last used program
4. Release the air pushing the RUN/STOP button
(temperature)
(duration)
(2ª temperature)
(End cycle)
(2ª duration)
(cooldown time)
prolonged
beep
air/water
release**
bars
minutes
As explained in the example with APOLIX you can set 2 temperatures (t and t1) and 2 polymerization times (du and du1):
t= first polymerization temperature
t1= second polymerization temperature (must always be at least 1° lower than t)
du= first polymerization time
du1= second polymerization time
CO= cooling time
SETTING OF THE PRESSURE
1. Turn on the machine using the ON/OFF switch (2)
2. Hold the two arrows (18) at the same time for 3 seconds (18). The display will say “AIR” (19)
3. The pressure will be the same as the last used program.
4. Release the air pushing the RUN/STOP button (17)
minutes