background image

 5 

Change the Settings in Logger 

Pro

 3 

Spectrometer Dialog Box 

The Spectrometer dialog box lists all the settings for the device. To display this box 
choose Set Up Sensors 

 Spectrometer from the Experiment menu. 

For most experiments, the default settings work well. 
There are four parameters listed in the dialog box. 

 

Sample Time: this is similar to the shutter speed of a camera. Logger 

Pro

 3 

automatically selects the proper sample time during calibration. 

Note:

 For 

emission studies, you may need to change the sample time manually. 

 

Wavelength Smoothing: the number of adjacent readings on either side of a given 
value that is used to calculate an average value. 

Note

: Be careful adjusting this 

parameter as it may shift your wavelength values slightly.  

 

Samples to Average: the number of readings taken at a given wavelength to 
calculate an average reading. 

 

Wavelength Range: the range is determined by the type of spectrometer in use. 

 

By clicking on the picture of the spectrometer in this dialog box, you will gain 
access to four options: calibrate, configure data collection, go to the support web 
page, and units of measure. Click on an item to select it. 
 

Use the Vernier Spectrometer with LabQuest 

Getting Started 

1. Ensure the proper LabQuest App software is installed on your LabQuest before 

using the Vernier Spectrometer. Version 2.0 or newer, is required for LabQuest 2 
and version 1.1, or newer, is required for the original LabQuest.  

2. Use the USB cable to connect the Spectrometer to LabQuest.  
3. Turn on LabQuest. The LabQuest App will launch automatically and the Meter 

screen will be displayed. 

 

Select the Type of Data (or Units) You Want to Measure 

The default data type is absorbance. If you want to measure the absorbance of a 
solution, proceed directly to the Calibrate section.  
If you want to measure %T or intensity, do the following: 
1. From the Sensors menu, choose Change Units 

 USB: Spectrometer. 

2. Select the unit or data type you wish to measure. 

 

Calibrate the Spectrometer (Not Required if Measuring Intensity) 

1. Choose Calibrate 

 USB: Spectrometer from the Sensors menu.

 Note

: For best 

results, allow the Spectrometer to warm up for a minimum of five minutes. 

2. Fill a cuvette about ¾ full with distilled water (or the solvent being used in the 

experiment) to serve as the blank. After the Spectrometer has warmed up, place 
the blank cuvette in the Spectrometer. Align the cuvette so a clear side of the 
cuvette is facing the light source.  

3. Follow the instructions in the dialog box to complete the calibration, and then 

click 

 

Measurement 

vs

. Wavelength (Generate a Spectrum) 

1. Fill a cuvette about ¾ full of the solution to be tested and place it in the 

Spectrometer. 

 6

2. Start data collection by tapping on the green Start button in the lower left corner 

of the screen. Tap the red Stop button to end data collection. 

3. Select wavelength. 

Note:

 The wavelength of maximum absorbance (

λ

 max) is 

automatically selected. This 

λ

 max will be used for any subsequent data 

collection, such as a Beer’s law experiment (abs 

vs.

 conc.) or a kinetics 

experiment (abs 

vs

. time). If you wish to choose another wavelength, you can tap 

on the graph to select a new wavelength. Another way to change the wavelength 
is to navigate to the Meter screen, tap on the meter itself, and select Change 
Wavelength. Enter the wavelength of your choice and select OK. If the 
wavelength you type in is not measured by the unit, the LabQuest will 
automatically choose the wavelength closest to your choice. 

4. To store the spectrum data, tap on the file cabinet icon in the upper right of your 

screen. 

 

Measurement 

vs

. Concentration (Beer’s law Studies) 

1. Generate a spectrum as described above. On the Meter screen, tap Mode. Change 

the mode to Events with Entry. 

2. Enter the Name (e.g., Concentration) and Units (e.g., mol/L). Select OK. 
3. A message will appear warning you to either save or discard the full spectrum run. 

Make your choice and proceed with data collection. 

4. Place your first Beer’s law standard solution in the Spectrometer. Start data 

collection. After the absorbance reading stabilizes, tap Keep. Enter the 
concentration of the solution and select OK. 

5. Place your second standard sample in the Spectrometer. After the absorbance 

readings stabilize, tap Keep. Enter the concentration of the second sample and 
select OK. 

6. Repeat Step 5 for the remaining standard samples. After you have tested the final 

standard, tap the red Stop button to end data collection. 

7. To calculate a best fit line equation for your standards, choose Curve Fit from the 

Analyze menu. Select Linear for the Fit Equation, and then select OK. The graph 
screen will appear again with the linear regression equation displayed. 

8. Place a cuvette containing an unknown sample of solution in the Spectrometer. 

Tap the Meter tab and write down the displayed absorbance value. Tap the Graph 
tab and choose Interpolate from the Analyze menu. Trace the linear regression 
equation to determine the concentration of the unknown. 

 

Measurement 

vs

. Time (Kinetics) 

1. Generate a spectrum as described above. On the Meter screen, tap Mode. Change 

the data-collection mode to Time Based. 

2. You can change the rate, interval, and/or duration of time of data collection, if 

desired. Select OK when you are ready to proceed. 

3. A message will appear warning you to either save or discard the full spectrum run. 

Make your choice and proceed with data collection. 

4. Mix the reactants, transfer ~2 mL of the reaction mixture to a cuvette and place 

the cuvette in the Spectrometer. Start data collection. You may tap the red Stop 
button to end data collection early. 

5. To calculate a function for your data, choose Curve Fit from the Analyze menu. 

Select the Fit Equation, and then select OK. The graph screen will appear again. 

 

Отзывы: