background image

 3 

can change the center wavelength value by clicking on the graph or by 
choosing a wavelength from the list. 

 

Option 2 

 If you wish to use the 

λ

 max chosen by Logger 

Pro

, but you want the 

absorbance to be measured 

only

 at that one wavelength, change Single 10 nm 

Band to Individual Wavelengths. You may then select up to ten wavelengths to 
measure at the same time.  

 

Option 3 

 If you wish to measure an average over a range of contiguous 

wavelengths of your choice, change Single 10 nm Band to Individual 
Wavelengths. Click 

. Select boxes in the list or drag your cursor on 

the graph to select up to ten contiguous wavelengths. Check Combine 
Contiguous Wavelengths.  

4. Click 

 to continue. 

5. Click 

. Place your first sample in the cuvette slot of the Spectrometer. 

After the readings stabilize, click 

. Enter the concentration of the sample 

and click 

6. Place your second sample in the cuvette slot. After the readings stabilize, click 

. Enter the concentration of the second sample and click 

7. Repeat Step 6 for the remaining samples. When finished, click 

 to end 

data collection. 

8. Click Linear Fit,  , to see the best fit line equation for the standard solutions. 
9. If doing Beer’s law to determine the concentration of an unknown, place the 

unknown sample in the cuvette holder. Choose Interpolation Calculator from the 
Analyze menu. A helper box will appear, displaying the absorbance and 
concentration of the unknown. Click 

 

Measurement 

vs

. Time (Kinetics) 

1. Generate a spectrum as described above. 
2. Click the Configure Spectrometer Data Collection button, 

.  

3. Select Absorbance 

vs

. Time as the data-collection mode. The wavelength of 

maximum absorbance will be selected. Click 

 to continue or click 

 and select a wavelength on the graph or in the list of wavelengths. 

See the previous section for more details. 

4. The default settings are 1 sample per second for 200 seconds. To change the  

data-collection parameters for your experiment, choose Data Collection from the 
Experiment menu and make the necessary changes. Click 

5. Mix the reactants. Transfer ~2 mL of the reaction mixture to a cuvette and place 

the cuvette in the Spectrometer. Click 

. Click 

 if you wish to end 

data collection early. 

6. Click Curve Fit,  , to calculate a function for your data.  

 

 4

Measure Emission Spectra with a Computer 

You may use your Spectrometer to measure the emission spectrum of a light source 
such as an LED or a gas discharge tube. To do so, you will need to purchase an 
optical fiber assembly (order code: VSP-EM-FIBER).  

Measure Intensity of Light Emissions 

1. Use a small screwdriver to loosen the two screws that connect the cuvette holder 

to the Spectrometer (see Figure 1). Remove the cuvette holder and connect an 
optical fiber assembly to the Spectrometer. 

 

 
 
 
 

Figure 1 

 

2. Use a USB cable to connect the Spectrometer to your computer. 
3. Start Logger 

Pro

 3. 

4. Choose Change Units 

 Spectrometer 

 Intensity from the Experiment menu. 

Intensity is a relative measure with a range of 0–1. 

Note:

 The Spectrometer is not 

calibrated for measuring intensity. 

5. Aim the tip of the optical fiber cable at a light source. Click 

. Click 

 to end data collection.  

 

If the spectrum maxes out (flat and wide peaks at a value of 1), increase the distance 
between the light source and the tip of the optical fiber cable or reduce the sample 
time (see Change the Settings in Logger 

Pro 

3 below).  

To increase the sample time, or if data collection is unusually slow, choose Set Up 
Sensors 

 Spectrometer: 1 from the Experiment menu. Set the Sample Time (begin 

with 75 ms, with subsequent reductions by 20 ms) to a suitable value and decrease 
the Samples to Average to 1.  
 

Use the Stored Emissions Files in Logger 

Pro

 3 

Logger 

Pro

 3 contains a folder of emissions graphs from selected discharge tubes, 

including: argon, helium, hydrogen, mercury, oxygen, sodium, and xenon. You can 
display and analyze these graphs without a Spectrometer connected to your 
computer. Follow the steps below to view one of these graphs. 
1. Choose Open from the File menu. 
2. Open the Sample Data folder. 
3. Inside the Sample Data folder, open the Physics folder. 
4. Inside the Physics folder, open the Gas Discharge Spectra. Open the desired file. 

 

You can use the mercury emissions graph to test fluorescent lighting for the presence 
of mercury. 

 

Отзывы: