background image

 

Vernier Spectrometer 

(Order Code: V-SPEC) 

 

The Vernier Spectrometer is a portable visible light 
spectrometer.  
 

What is Included with the Vernier Spectrometer? 

 

Spectrometer (with light source/cuvette holder attached) 

 

One package of 15 plastic cuvettes and lids 

 

USB cable 

 

Software Requirements 

Logger 

Pro

®

 3 (version 3.8.5 or newer) software is required if you are using a 

computer. LabQuest App version 2.0, or newer, is required if you are using 
LabQuest

®

 2. LabQuest App 1.1, or newer, is required if you are using the original 

LabQuest. Visit the downloads section of www.vernier.com to update your software. 

NOTE: 

Vernier products are designed for educational use. Our products are not 

designed nor are they recommended for any industrial, medical, or commercial 
process such as life support, patient diagnosis, control of a manufacturing process, or 
industrial testing of any kind.  

 

Safety 

 

When this instrument is in data-collection mode labeled “Intensity”, the light 
source will be blocked or turned off. Continue to use proper safety precautions.  

 

Do not remove or modify any of the installed safety components of this device. 
Doing so will create an unsafe operating condition and will void the product 
warranty. 

 

No user-serviceable parts are in this device. Do not attempt to open or modify this 
device. Contact Vernier for all repairs and service including lamp replacement. 

 

Handle the device with care. This instrument can be damaged if it is dropped.  

 

Do not use this instrument if it is damaged in any way. Contact Vernier Technical 
Support for troubleshooting and technical assistance. 

 

Do not use this instrument for clinical or diagnostic procedures.  

 

Use the Vernier Spectrometer with a Computer 

(For LabQuest instructions, see page 5) 

 

Getting Started 

1. Ensure Logger 

Pro

 software (version 3.8.5 or newer) is installed on your 

computer before using the Vernier Spectrometer. 

2. Connect the Spectrometer to a powered USB port on the computer or a powered 

USB hub. 

Note:

 The first time you connect a spectrometer, your computer may 

ask you a few questions. 

Do not go online for device drivers

. The device drivers 

were installed when you installed Logger 

Pro

 3. 

3. Start Logger 

Pro

 software on your computer. 

 

 2

Select the Type of Data (or Units) You Want to Measure 

The default data type is absorbance. If you want to measure the absorbance of a 
solution, proceed directly to the Calibrate section below.  
If you want to measure %T or intensity, do the following: 
1. Choose Change Units 

 Spectrometer from the Experiment menu. 

2. Select the unit or data type you wish to measure. 

 

Calibrate (Not Required if Measuring Intensity) 

1. To calibrate the Spectrometer, choose Calibrate 

 Spectrometer from the 

Experiment menu.

 Note

: For best results, allow the Spectrometer to warm up for a 

minimum of five minutes. 

2. Fill a cuvette about ¾ full with distilled water (or the solvent being used in the 

experiment) to serve as the blank. After the Spectrometer has warmed up, place 
the blank cuvette in the Spectrometer. Align the cuvette so the clear side of the 
cuvette is facing the light source.  

3. Follow the instructions in the dialog box to complete the calibration, and then 

click 

 

Collect Data with a Computer 

There are three general types of data collection that measure absorbance or 
transmittance—absorbance (or %T) 

vs.

 wavelength, which produces a spectrum, 

absorbance (or %T) 

vs

. concentration for Beer’s law experiments, and absorbance 

(or %T) 

vs. 

time for kinetics experiments. 

Measurement 

vs

. Wavelength (Generate a Spectrum) 

1. Fill a cuvette about ¾ full of a sample of the solution to be tested. Place the 

sample in the Spectrometer and click 

. Click 

 to end data 

collection. 

2. To store the spectrum data, choose Store Latest Run from the Experiment menu. 

 

Measurement 

vs

. Concentration (Beer’s law Studies) 

1. Generate a spectrum as described above. 
2. Click the Configure Spectrometer Data Collection button, 

  There are three regions in this box: 

 

Collection Mode

  The three options for data collection are offered. If the 

measurement (Absorbance in this example) 

vs

. Time or 

vs

. Concentration is 

selected, a wavelength or wavelengths will need to be chosen.  

 

Graph 

 The graph displays a full-spectrum analysis of the sample in the 

cuvette holder. By default, the wavelength with the maximum measured value 
will be selected. You may wish to select a different wavelength. See Step 3 for 
details. 

 

List of wavelength options

  This column lists all the available wavelengths. It 

becomes active when either the Concentration or Time mode is selected. 

3. Select Absorbance (or %T) 

vs

. Concentration as the data-collection mode. The 

wavelength with the maximum value from the spectrum (

λ

 max) will be 

automatically selected. There are three options when choosing a wavelength (or 
wavelengths) for subsequent measurements.  

 

Option 1

  The default option is to use a single 10 nm band. This measures the 

average absorbance from ~5 nm on either side of the chosen wavelength. You 

Отзывы: