Specifications

Pressure

Type: Differential

Range: ±500 Pa

Accuracy: ±1 Pa or 3% of reading

Resolution: 0.02 Pa

Flow rate range

±10 L/s

Flow head

Volume: 41 mL

Dimensions: 29.5 mm (diameter) × 82 mm
(length)

Mass: 69 g

Construction: ABS plastic

Adjusted volume calculation

Returns volume to zero after each ventilation
cycle

Cycle volume calculation

Reports the peak volume for each ventilation
cycle

Respiration rate calculation

Sample Window: 30 s

Advance Window: 10 s

Wireless specification

Bluetooth 4.2

Maximum wireless range

30 m

Battery

650 mA Li-Poly

Battery life (single full
charge)

~24 hours

Battery life (long term)

~500 full charge cycles (several years
depending on usage)

Normal operating
temperature range

20ºC to 40ºC

Operating humidity range

5–95% (non-condensing)

Care and Maintenance

Every person must use their own bacterial filter and mouthpiece—do not share
these.

The flow head is not designed to be removed or sterilized by the user. The flow
head can be replaced to extend the life of the product. For more information,
contact Vernier Technical Support at [email protected] or call 888-837-
6437.

Differential Pressure

This channel measures the pressure differential (Pa) across the screen inside the
flow head. This channel is not active by default when the sensor is connected.

Respiration Rate

This channel detects inhalations and calculates the number of breaths per minute
(BPM). The sample window for the calculation is 30 seconds. The advance
interval is 10 seconds. The value will update every 10 seconds. Use this channel
for long experiments. This channel is not active by default when the sensor is
connected.

Calibrating the Sensor

Go Direct Spirometer is factory calibrated and should not require calibration by
the user.

For the most accurate measurements, we recommend adjusting the volume
channel using a 2 L syringe (not included). This is a simple process that takes
only a few minutes.

For further calibration information, see

www.vernier.com/til/3353

Содержание Go Direct Spirometer

Страница 1: ...bQuest 2 You are now ready to collect data 4 This is a multi channel sensor To change the channel selections see www vernier com start gdx spr Charging the Sensor Connect Go Direct Spirometer to the included Micro USB Cable and any USB device for two hours You can also charge up to eight Go Direct Spirometers using our Go Direct Charge Station sold separately order code GDX CRG An LED on each Go D...

Страница 2: ...y default when the sensor is connected Charging Blue LED on steady while sensor is connected to the Micro USB Cable or Charging Station Fully charged Blue LED is off when charging is complete Powering the Sensor Turning on the sensor Press button once Red LED indicator flashes when unit is on Putting the sensor in sleep mode Press and hold button for more than three seconds to put into sleep mode ...

Страница 3: ...is not designed to be removed or sterilized by the user The flow head can be replaced to extend the life of the product For more information contact Vernier Technical Support at support vernier com or call 888 837 6437 Differential Pressure This channel measures the pressure differential Pa across the screen inside the flow head This channel is not active by default when the sensor is connected Re...

Страница 4: ...adio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures The nose clips should be wiped clean between uses and soaked in a mild detergent after each experiment Battery Information Go Direct Spirometer contains a small lithium ion battery The system is designed to consu...

Страница 5: ...his device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device To reduce potential radio interference to other users the antenna type and its gain should be so chosen that the equivalent isotropically radiated power e i r p is not more than that permitted for successful communication RF exposure warning T...

Отзывы: