background image

7

Specifications

Detection modes

Absorbance, % Transmittance, Fluorescence,
Raw Lamp Output, and Intensity (fiber
required)

Dimensions

18.5 cm × 17 cm × 7 cm

Lamp power supply

AC adapter (included)

Power consumption

3 A start-up, 500 mA continuous

Bluetooth power supply

USB (USB adapter included)

Absorbance light source

Deuterium (UV) and Incandescent (VIS)

Detector

Linear CCD

Emission light source

Exchangeable LED included (375 nm,
450 nm, 525 nm)

Wavelength range

220 nm–850 nm

Wavelength reporting interval

~1 nm

Optical resolution

3.0 nm (as determined with 486 nm
hydrogen emission spectral line FWHM)

Wavelength accuracy

± 2.0 nm (as determined with holmium
oxide NIST standard)

Absorbance photometric accuracy ± 5.0% (as determined with potassium

dichromate NIST standards)

Typical scan time

~2 s

Sample format

10 mm × 10 mm cuvette (UV fluorescence
cuvette included)

Fluorescence emission detection
limit

1 mg/L quinine sulfate dihydrate in 0.1 M
H

2

SO

4

Beam height (z dimension)

8.5 mm

Safety

l

Always wear UV blocking safety glasses or goggles when using this device.
The light source emits ultraviolet radiation and can cause eye damage.

l

When this instrument is in data-collection mode labeled Intensity, the light
source will be blocked or turned off. Continue to use proper safety precautions.

l

This instrument contains an internal high-voltage power source. Position the
instrument where the ON/OFF switch can be easily reached. Do not move the
instrument when it is running. Do not attempt to open or remove the case at
any time.

If the spectrum maxes out (flat and wide peaks at a value of 1), increase the
distance between the light source and the tip of the optical fiber cable or reduce
the sample time (see Change the Settings in Logger

Pro

).

To increase the sample time, or if data collection is unusually slow, choose Set Up
Sensors ► Spectrophotometer: 1 from the Experiment menu. Set the Sample Time
(begin with 75 ms, with subsequent reductions by 20 ms) to a suitable value and
decrease the Samples to Average to 1.

Use the Stored Emissions Files in Logger

Pro

Logger

Pro

contains a folder of emissions graphs from selected discharge tubes,

including: argon, helium, hydrogen, mercury, oxygen, sodium, and xenon. You
can display and analyze these graphs without a spectrometer connected to your
computer. Follow these steps to view one of these graphs.

1. Choose Open from the File menu.
2. Open the Sample Data folder.
3. Inside the Sample Data folder, open the Physics folder.
4. Inside the Physics folder, open the Gas Discharge Spectra. Open the desired

file.

You can use the mercury emissions graph to test fluorescent lighting for the
presence of mercury.

Change the Settings in Logger

Pro

The Spectrophotometer dialog box lists all the settings for the device. To display
this box choose Set Up Sensors ► Spectrophotometer from the Experiment menu.

For most experiments, the default settings work well.

There are five parameters listed in the dialog box.

l

Sample Time: This is similar to the shutter speed of a camera. Logger

Pro

automatically selects the proper sample time during calibration.

Note:

For

emission studies, you may need to change the sample time manually.

l

Wavelength Smoothing: This is the number of adjacent readings on either side
of a given value that is used to calculate an average value.

Note:

Be careful

adjusting this parameter as it may shift your wavelength values slightly.

l

Samples to Average: This is the number of readings taken at a given
wavelength to calculate an average reading.

l

Wavelength Range: The range is determined by the type of Spectrophotometer
in use.

l

LED Intensity: This allows you to change the intensity of the excitation LED.

By clicking on the picture of the Spectrophotometer in this dialog box, you will
gain access to four options: calibrate, configure data collection, go to the support
website, and units of measure. Click on an item to select it.

Videos

View videos related to this product at

www.vernier.com/gdx-spec-fuv

Содержание Go Direct GDX-SPEC-FUV

Страница 1: ...etooth icon is solid when sensor is connected via Bluetooth wireless tech nology Go Direct Fluorescence UV VIS Spectrophotometer Order Code GDX SPEC FUV Go Direct Fluorescence UV VIS Spectrophotometer...

Страница 2: ...discharge tube To do so you will need to purchase the Vernier Optical Fiber order code VSP FIBER Measure Intensity of Light Emissions Insert the Vernier Optical Fiber into the spectrometer Connecting...

Страница 3: ...of your screen 1 Connect the spectrometer following the steps in the Getting Started section of this user manual 2 Open Spectral Analysis 3 Select the appropriate Emissions experiment from the listed...

Страница 4: ...he Meter screen tap Mode Change the mode to Events with Entry 2 Enter the Name e g Concentration and Units e g mol L Select OK 3 A message will appear warning you to either save or discard the full sp...

Страница 5: ...ments the default settings work well There are five parameters listed in the dialog box l Sample Time This is similar to the shutter speed of a camera LabQuest automatically selects the proper sample...

Страница 6: ...at a light source Click Click to end data collection Note The Spectrophotometer is not calibrated for measuring intensity l Option 1 The default option is to use a single 10 nm band This measures the...

Страница 7: ...e Sample Time begin with 75 ms with subsequent reductions by 20 ms to a suitable value and decrease the Samples to Average to 1 Use the Stored Emissions Files in Logger Pro Logger Pro contains a folde...

Страница 8: ...this transmitter must not be co located or operating in conjunction with any other antenna or transmitter Le pr sent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio e...

Страница 9: ...ensitive measurement device For best results use the cables that were provided Keep the device away from electromagnetic noise sources such as microwaves monitors electric motors and appliances Vernie...

Отзывы: