5.  Re-assemble the tester.

Specifications

Output Voltage: 2.5 volts DC @ 10mA
Dimensions: 12 cm x 4.5 cm x 2 cm
Operating Temperature: 0° to 40°C

Calibration

The Low Resistance Tester is calibrated to standards 

traceable to NIST.  Frequency of recalibratrion should be 

based on the critical nature of those ESD sensitive items 

handled and the risk of failure for the ESD protective 

equipment and materials.  In general, we recommend that 

calibration be performed annually.
Calibration is performed by placing various resistors between 

the probe tip and the lead wire.
1.  Attach the alligator clip to the lead wire.
2.  Set the Low Resistance Tester to the 1 ohm test setting.
3.  Attach the lead of a 1 ohm resistor to the alligator clip.
4.  Firmly place the tester’s probe tip on the opposite resistor 

 lead.
5.  Press and hold the test button.  The PASS LED should 

 illuminate.
6.  Repeat the procedure with a 1.2 ohm resistor.  The PASS 

  LED should not illuminate.
7.  Repeat this process with the ranges listed in the table 

  below.  If the unit fails calibration, check and replace the 

 battery.

Test Range Setting

PASS

FAIL

1 ohm

0 to 1 ohm

≥ 1.2 ohms

2 ohms

0 to 2 ohms

≥ 2.2 ohms

20 ohms

0 to 20 ohms

≥ 22 ohms

Maintenance

BATTERY REPLACEMENT

1.  Properly ground yourself using a wrist strap.
2.  Remove the 2 screws located at the back of the tester.

Figure 5.  Locating the enclosure screws

Figure 6.  Removing the back cover

3.  Remove and turn over the circuit board.
4.  Locate and replace the battery (3 Volt; Model CR2032).

First, place 

battery under 

prongs

Next, press here to snap 

battery in place

Figure 7.  Replacing the battery