5. Press “
CHA
+” button again to change to work time adjustment screen.
Flashing
WORK
segment indicates that work period can be adjusted.
Use “
CHB +”
or “
CHB - “
button to adjust the time in range from 1 � 99
seconds (1 sec. increments).
6. Press “
CHA
+” button again to change to rest time adjustment screen.
Flashing
REST
segment indicates that rest period can be adjusted. Use
“
CHB +”
or “
CHB - “
button to adjust the time in range from 1 � 99 sec�
onds (1 sec. increments).
7. Press “
CHA
+” button again to change to ramp up time adjustment
screen. Flashing
RMP
segment indicates that ramp up time can be ad�
justed can be adjusted. Use “
CHB +”
or “
CHB - “
button to adjust the time
in range 0.3 � 9.9 seconds ( 0.1 sec increments)
8. Press “
CHA
+” button again to change to ramp up time adjustment
screen. Flashing
RMP
segment indicates that ramp up time can be ad�
justed can be adjusted. Use “
CHB +”
or “
CHB - “
button to adjust the time
in range 0.3 � 9.9 seconds ( 0.1 sec increments)
9. Press “
CHA
+” button again to change to output SYN/ALT selection screen
(synchronous or alternating) . Flashing
SYN (or ALT)
symbol
indicates
that output type can be set to synchronous (SYN) or alternating (ALT).
Use “
CHB +”
or “
CHB - “
button to select SYN or ALT.
Synchronous (SYN)
means that both Channel A and Channel B will give
output at the same time of each Work period.
Alternating (ALT)
means that Channel A will work while Channel B will
be off for one Work period and then for the next Work period Channel B
will be on while but Channel A will be off and so on until the end of the
programme time.
10. If SYN (Synchronous) has been selected, press the “
CHA
+” button
again to change to delay time adjustment screen. Flashing
DLY
symbol
indicates that delay time of channel B can be adjusted. Use “
CHB +”
or
“
CHB - “
button to set required time in range 0 � 4 seconds ( 0.1 sec incre�
ments).
11. Once delay ( or output synchronization) has been adjusted press SET/ESC
button to save settings and return to Home screen.
If “
CHA
+” or “
CHB
�” button was pressed again rather than SET/ESC then
the device would loop back to the Hz adjustment screen .
12. Repeat above procedure to readjust custom programme.