7.8
Solid-State Relay Replacement
Disconnect the product from the power supply and remove the
appropriate cover as given above.
Make a note of the wire connections to the solid state relay and disconnect them.
Remove the solid state relay from the base panel or aluminium plate.
Replace and reconnect the solid state relay ensuring that the bottom of it has good
thermal contact with the base panel or aluminium plate.
Replace the access panel.
33
7.0 Repairs and Replacements
Содержание CARBOLITE GERO VMF-ASTM
Страница 4: ...10 2 Power Settings 38 11 0 Specifications 39 11 1 Environment 39 4 ...
Страница 20: ...3 0 2132 Controller 3 7 Navigation Diagram 20 ...
Страница 28: ...6 0 Maintenance Power Consumption Measure the current drawn on each phase cir cuit 28 ...
Страница 37: ...9 2 WA 11 01 Connections below show single phase with over temperature controller 37 9 0 Wiring Diagrams ...
Страница 40: ......
Страница 41: ...Notes Service Record Engineer Name Date Record of Work ...