7.0 Maintenance
7.7
Power Adjustment (Thyristor)
The current-limiting thyristor stacks which control power to the elements are fitted with
an adjustable resistor which is factory set to limit the maximum current supplied. In the
event of a change of supply voltage, or the fitting of a new thyristor, further adjustment
may be required.
The maximum element currents for this model are listed in section 11.0. Please contact
Carbolite Gero for further information.