2.3
Fitting the Heating Elements
Please refer to the Element Installation and Replacement in section 8.0.
Wear eye protection when handling the heating elements.
The molybdenum disilicide elements are EXCEPTIONALLY FRAGILE and
are packed separately, together with other items shown in the list.
items packed
quantity
Elements
6
Element Clamps
12
Element Clips
12
Braids
1 set
Separators/ Block
6
Chimney Unit
1
Remove the top access panel to allow element fitting.
Fit the heating elements vertically into the chamber. Handle the elements with great
care. The installation of these elements is described in section 8.9. Connect the
aluminium braids securely as indicated in the element installation instructions.
2.4
Electrical Connections
Connection by a qualified electrician is recommended.
The product covered by this manual normally requires a single phase A.C. supply, which
may be Live to Neutral non-reversible, Live to Neutral reversible or Live to Live. Some
models may be ordered for 3-phase use, which may be star or delta.
Check the product rating label before connection. The supply voltage should agree with
the voltage on the label and the supply capacity should be sufficient for the current on
the label.
The supply should be fused at the next size equal to, or higher than the current on the
label. A table of the most common fuse ratings is also given towards the back of this
manual. When the mains cable is factory fitted, internal fuses are also fitted. It is
essential that the operator ensures that the product is correctly fused.
Products with a factory fitted supply cable are designed to be wired directly to an
isolator or fitted with a line plug.
Products without a factory fitted supply cable require a permanent connection to a
fused and isolated supply. The product's electrical access panel should be temporarily
removed, and connections made to the internal terminals.
If the product is to be connected by line plug. The plug should be within reach of the
operator and should be easy to remove.
9
2.0 Installation
Содержание CARBOLITE GERO HTF 18/15
Страница 36: ...4 0 3508 Dual Loop Cascade Power Limit View 36 ...
Страница 55: ...Element Connections 1 phase Element Connections 3 phase 55 8 0 Repairs and Replacements ...
Страница 58: ...10 0 Wiring Diagrams 10 0 Wiring Diagrams 10 1 239 4 5001 58 ...
Страница 61: ...Notes Service Record Engineer Name Date Record of Work ...