11.0
Fuses and Power Settings
11.1
Fuses
F1-F2: Refer to the circuit diagrams.
F1
Internal
Supply
Fuses
Fitted if supply cable fitted.
Fitted on board to some types
of EMC filter.
GEC Safeclip of the type shown
(glass type F up to 16 A)
38 mm x 10 mm type F fitted on
EMC filter circuit board(s)
F2
Auxiliary
Circuit Fuses
Fitted on board to some types
of EMC filter.
May be omitted up to 25 Amp/
phase supply rating.
2 Amps glass type F
On board: 20 mm x 5 mm
Other: 32 mm x 6 mm
Thyristor Fuse
Ferraz Protistor of the rating
shown
Customer
Fuses
Required if no supply cable
fitted.
Recommended if cable fitted.
See rating label for current;
See table below for fuse rating.
Model
Phases
Volts
Supply Fuse Rating
Type
HTF 18/15
1-phase
220-240
50 A
ES50
HTF 18/15
3-phase+N
380/220 - 415/240
16 A
type F1
HTF 18/15
3-phase delta
220-240
32 A
NS32
11.2
Power Settings
The power limit settings (parameter OP.Hi) for this model are voltage dependant. The
figures represent the maximum percentage of time that controlled power is supplied to
the elements. Do not attempt to “improve performance” by setting a value higher than
the recommended values. To adjust the parameter refer to the "Changing the
Maximum Output Power" of the control section of the manual.
Volts (V)
208
220
230
240
380
400
415
Power (%)
100
100
100
100
100
100
100
Please refer to the rating label for product specific information.
59
11.0 Fuses and Power Settings
Содержание CARBOLITE GERO HTF 18/15
Страница 36: ...4 0 3508 Dual Loop Cascade Power Limit View 36 ...
Страница 55: ...Element Connections 1 phase Element Connections 3 phase 55 8 0 Repairs and Replacements ...
Страница 58: ...10 0 Wiring Diagrams 10 0 Wiring Diagrams 10 1 239 4 5001 58 ...
Страница 61: ...Notes Service Record Engineer Name Date Record of Work ...