10.0 Fuses and Power Settings
phase supply rating.
Customer
Fuses
Required if no supply cable
fitted.
Recommended if cable fitted.
See rating label for current;
See table below for fuse rating.
Model Phases
Volts
Supply Fuse Rating
4/28
1-phase
220-240
6 A
5/18
1-phase
230
10 A
6/500 3-phase + N
380-415
25 A
10.4
Power Settings
The power limit settings (parameter OP.Hi) for this model are voltage dependant. The
figures represent the maximum percentage of time that controlled power is supplied to
the elements. Do not attempt to “improve performance” by setting a value higher than
the recommended values. To adjust the parameter refer to the "Changing the
Maximum Output Power" of the control section of the manual.
Volts
208 V
220 V
230 V
240 V
380 V
400 V
415 V
Power (%)
88
100
100
100
-
-
-
Power (%)
-
-
100
-
-
-
-
Please refer to the rating label for product specific information.
26
Содержание CARBOLITE GERO HTCR 4/28
Страница 5: ...Caution Double Pole Neutral Fusing 5 1 0 Symbols and Warnings ...
Страница 28: ......