background image

OPERATING MANUAL

DRYING OVEN

NB-901M

N-BIOTEK, INC.
402-803 Technopark, 655, Pyeongcheon-ro, Bucheon-si,
Gyeonggi-do, 14502, Korea
Tel.
Fax.
Web.
Email.

82.32.321.2100
82.32.328.2372

http://www.n-biotek.com

[email protected]

Содержание NB-901M

Страница 1: ...ING MANUAL DRYING OVEN NB 901M N BIOTEK INC 402 803 Technopark 655 Pyeongcheon ro Bucheon si Gyeonggi do 14502 Korea Tel Fax Web Email 82 32 321 2100 82 32 328 2372 http www n biotek com export n biot...

Страница 2: ...ot make the power plug be pressed by back of the product A space between the product and the plug must be 20cm at least The power outlet must be only for this product Using various products simultaneo...

Страница 3: ...flat it can cause an excessive vibration of the product When you move the product do not lay down to its side or reverse the head to bottom It can cause a malfunction Open and close the door softly a...

Страница 4: ...of pressure by heat inside leave the air vent configured in the device open and use it It can cause burns This product is generally used at high temperatures Be sure to wear safety gloves that can wit...

Страница 5: ...igned Digital P I D Controller Designed to put the test object in multiple stages by installing a shelf SUS WIRE inside the chamber Minimize the risk of internal fire with PIN HEATER By installing the...

Страница 6: ...6 N BIOTEK l NB 901M Ver 2021 3 1 Configuration 3 Configuration and Control Panel 3 2 Control Panel FRONT BACK DOOR HAND SIGHT GLASS FAN MOTOR CONTROL PANEL RUBBER FEET 1 2 3 4 5 6 7 8 9 10 11 12...

Страница 7: ...icate heater 5 Timer Lamp Indicate timer 6 Auto Tuning Lamp Indicate auto tuning run 7 Start Stop Button On Off equipment 8 Parameter Setting Button Parameter setting 9 Shift L Button Movement of the...

Страница 8: ...o change setting value you should release after push SET button TEMP sign will appear on the Temp display Press SET button one more time the first digit will blink waiting for input Set your wanted te...

Страница 9: ...not working until get to the setting temperature as flickering the display of timer The unit will be started when current temp gets to setting temp display of time will stop flickering and timer start...

Страница 10: ...NG make to stop with AT button over 5 sec Display a setting temperature as following for AUTO TUNING 4 6 Parameter setting Please do as following for the parameter setting 1 Press SET button over 5 se...

Страница 11: ...timer Absolute value Present fixing temperature ActP Timer start working Present temperature fixing temperature ACTP value Timer start working Present temperature fixing temperature ACTP value Timer s...

Страница 12: ...g mode control 7 Mode 1 Running mode control 8 Mode 2 Running mode control Type A timer end buzzer on Type B timer end heater off after AOFF value Alarm off display END beep Type C timer end heater of...

Страница 13: ...beginning power on Setting range 0 9999 min It just can use when cooler working delay time with mode N3 1 12 dTon Fixing temperature time for defrost ability on the screen Setting range 0 9999 min Fix...

Страница 14: ...eas of defective material and workmanship provided such defect results from normal and proper use of the equipment 1 The free warranty service will be provided once the unit is proved to be defective...

Отзывы: